The P6019 and P6020 split-core AC current probes were introduced in 1966 together with passive terminations and the Type 134 amplifier.

The probes had no internal frequency compensation components, contrary to their 1969 successor types P6021 and P6022.

Specifications

P6019 passive P6019 + 134 P6020 passive P6020 + 134
Sensitivity 2 mA/mV or 10 mA/mV 1 mA/Div to 1 A/Div (scope @ 50 mV/Div) 1 mA/mV or 10 mA/mV 1 mA/Div to 1 A/Div (scope @ 50 mV/Div)
Bandwidth (-3 dB) 60 MHz @ 10 mA/mV > 40 MHz 200 MHz @ 10 mA/mV > 70 MHz
LF limit (-3 dB) 450 Hz @ 2 mA/mV, 120 Hz @ 10 mA/mV < 12 Hz 6500 Hz @ 1 mA/mV, 775 Hz @ 10 mA/mV < 50 Hz
Maximum current 15 Ap-p 6 Ap-p