[[File:P6022+term.jpg|right|thumb|350px|P6022, 6 A / 200 MHz current probe]]
[[File:P6022+term.jpg|right|thumb|300px|P6022, 6 A / 200 MHz current probe]]
The '''Tektronix P6022''' is a 50-turn AC current probe for oscilloscopes.  It can be used with the included passive termination box (011-0106-00), a [[134|type 134 current probe amplifier]], or a [[3A9]] or [[7A14]] current probe amplifier scope plug-in.  Bandwidth is at least 200 MHz with the passive termination.   
The '''Tektronix P6022''' is a 50-turn AC current probe for oscilloscopes.  It can be used with the included passive termination box (011-0106-00), a [[134|type 134 current probe amplifier]], or a [[3A9]] or [[7A14]] current probe amplifier scope plug-in.   
[[File:P6022 circuit.jpg|right|thumb|350px|Schematic (click to enlarge)]]
 
Bandwidth is at least 200 MHz with the passive termination.   
 
The sister model [[P6021]] has lower bandwidth but also a lower LF cut-off frequency and higher peak current rating.
The sister model [[P6021]] has lower bandwidth but also a lower LF cut-off frequency and higher peak current rating.


Line 9: Line 11:
! P6021 passive
! P6021 passive
! P6021 + [[134]]
! P6021 + [[134]]
! P6021 + [[7A14]]
! P6022 passive
! P6022 passive
! P6022 + [[134]]
! P6022 + [[134]]
! P6022 + [[7A14]]
|-
|-
! Sensitivity
! Sensitivity
| 2 mA/mV or 10 mA/mV  || 1 mA/Div to 1 A/Div<br/>(scope @ 50 mV/Div)
| 2 mA/mV or 10 mA/mV  || 1 mA/Div to 1 A/Div<br/>(scope @ 50 mV/Div) || 1 mA/Div to 1 A/Div
| 1 mA/mV or 10 mA/mV  || 1 mA/Div to 1 A/Div<br/>(scope @ 50 mV/Div)
| 1 mA/mV or 10 mA/mV  || 1 mA/Div to 1 A/Div<br/>(scope @ 50 mV/Div) || 1 mA/Div to 1 A/Div
|-
|-
! Bandwidth (-3 dB)
! Bandwidth (-3 dB)
| 60 MHz @ 10 mA/mV    ||  > 40 MHz
| 60 MHz @ 10 mA/mV    ||  > 40 MHz ||  50 MHz
| 200 MHz @ 10 mA/mV  ||  > 70 MHz
| 200 MHz @ 10 mA/mV  ||  > 70 MHz ||  105 MHz
|-
|-
! LF limit (-3 dB)
! LF limit (-3 dB)
| 450 Hz @ 2 mA/mV<br/>120 Hz @ 10 mA/mV  ||  < 12 Hz
| 450 Hz @ 2 mA/mV<br/>120 Hz @ 10 mA/mV  ||  < 12 Hz ||  25 Hz
| 8500 Hz @ 1 mA/mV<br/>935 Hz @ 10 mA/mV  ||  < 100 Hz
| 8500 Hz @ 1 mA/mV<br/>935 Hz @ 10 mA/mV  ||  < 100 Hz ||  250 Hz
|-
|-
! Maximum current
! Maximum current
| colspan="2" align="center" | 15 A<sub>p-p</sub>
| colspan="3" align="center" | 15 A<sub>p-p</sub>
| colspan="2" align="center" | 6 A<sub>p-p</sub><br>(> 10 kHz @ 1 mV/mA, > 3 kHz @ 10 mV/mA, < 10 MHz)
| colspan="3" align="center" | 6 A<sub>p-p</sub><br>(> 10 kHz @ 1 mV/mA, > 3 kHz @ 10 mV/mA, < 10 MHz)
|}
|}


Line 38: Line 42:
P6022+term.jpg | P6022 current probe with passive terminator 011-0106-00
P6022+term.jpg | P6022 current probe with passive terminator 011-0106-00
P6022.jpg
P6022.jpg
P6022 circuit.jpg| Schematic (click to enlarge)
</gallery>
</gallery>


Comparison of P6022 with passive termination @ 10 mA/Div (top trace) vs. P6021+134 (bottom trace).
Comparison of P6022 with passive termination @ 10 mA/Div (top trace) vs. P6021+134 (bottom trace).

Revision as of 01:29, 27 December 2017

P6022, 6 A / 200 MHz current probe

The Tektronix P6022 is a 50-turn AC current probe for oscilloscopes. It can be used with the included passive termination box (011-0106-00), a type 134 current probe amplifier, or a 3A9 or 7A14 current probe amplifier scope plug-in.

Bandwidth is at least 200 MHz with the passive termination.

The sister model P6021 has lower bandwidth but also a lower LF cut-off frequency and higher peak current rating.

Specifications

P6021 passive P6021 + 134 P6021 + 7A14 P6022 passive P6022 + 134 P6022 + 7A14
Sensitivity 2 mA/mV or 10 mA/mV 1 mA/Div to 1 A/Div
(scope @ 50 mV/Div)
1 mA/Div to 1 A/Div 1 mA/mV or 10 mA/mV 1 mA/Div to 1 A/Div
(scope @ 50 mV/Div)
1 mA/Div to 1 A/Div
Bandwidth (-3 dB) 60 MHz @ 10 mA/mV > 40 MHz 50 MHz 200 MHz @ 10 mA/mV > 70 MHz 105 MHz
LF limit (-3 dB) 450 Hz @ 2 mA/mV
120 Hz @ 10 mA/mV
< 12 Hz 25 Hz 8500 Hz @ 1 mA/mV
935 Hz @ 10 mA/mV
< 100 Hz 250 Hz
Maximum current 15 Ap-p 6 Ap-p
(> 10 kHz @ 1 mV/mA, > 3 kHz @ 10 mV/mA, < 10 MHz)

Manuals

Pictures


Comparison of P6022 with passive termination @ 10 mA/Div (top trace) vs. P6021+134 (bottom trace). Signal is a sawtooth from a HP3325A @ 5 Vp-p into 50 Ω = 100 mAp-p.