The '''Tektronix P6106''' is a passive 250 MHz ×10 probe for 1 MΩ inputs.  It was available with 1, 2 or 3 meters of probe cable.  The 6106 has a coding pin for scope readout ×10.
The '''Tektronix P6106''' is a passive 250 MHz ×10 probe for 1 MΩ inputs.  It was available with 1, 2 or 3 meters of probe cable.  The 6106 has a coding pin for scope readout ×10.  Color coding tells, how long is their cable.  Orange means 2m,  Blue means 1m.


{{BeginSpecs}}
{{BeginSpecs}}
Line 19: Line 19:
* [[Media:P6106.pdf|Tek P6106 manual (PDF)]]
* [[Media:P6106.pdf|Tek P6106 manual (PDF)]]
* [[Media:070-5527-00.pdf|Tek P6106A manual (PDF)]]
* [[Media:070-5527-00.pdf|Tek P6106A manual (PDF)]]
==Pictures==
<gallery>
P6106_1m.JPG    | 1m variant, blue
P6106_2m.JPG    | 2m variant, orange
</gallery>





Revision as of 09:23, 17 December 2018

The Tektronix P6106 is a passive 250 MHz ×10 probe for 1 MΩ inputs. It was available with 1, 2 or 3 meters of probe cable. The 6106 has a coding pin for scope readout ×10. Color coding tells, how long is their cable. Orange means 2m, Blue means 1m.

Key Specifications

Attenuation 10 × ±3% with scope input at 1 MΩ ±2%
Compensation range 15 pF to 24 pF
Bandwidth >250 MHz (>150 MHz for 3 meter cable). Using the 1 meter probe with a scope having at least 325 MHz bandwidth results in a system bandwidth of 300 MHz typ.
Input capacitance 10.5 pF (1 m), 13 pF (2 m), 15.5 pF (3 m)