TM5006: Difference between revisions
No edit summary |
No edit summary |
||
Line 19: | Line 19: | ||
The rightmost compartment (slot 6) can supply [[TM500 high-power compartments|high-powered plugins]]. | The rightmost compartment (slot 6) can supply [[TM500 high-power compartments|high-powered plugins]]. | ||
The DC rails to the plugins are provided by a switchmode PSU. The supply is not universal input voltage | The DC rails to the plugins are provided by a switchmode PSU. The supply is not universal input voltage, the input voltage must be set with two internal jumpers. DC rails are 8 V, +26 V, and −26 V. The case has a cooling fan which is connected to the primary of the transformer. | ||
Available options include 02 Rear Inputs and 10 a set of 19-inch rackmount slides. | Available options include Opt.02, Rear Inputs, and Opt.10, a set of 19-inch rackmount slides. | ||
For option 02 six | For option 02, six BNC connectors and a 50-pin female [[D-sub connector|DD-50 D-sub connector]] are installed on the rear panel and pin headers soldered internally, but no internal wiring is pre-installed. | ||
The rack-mount kit has different sides compared to the standard model. It is not possible remove the standard carry handle to install a rack-mount kit. The entire side panel must be swapped. | The rack-mount kit has different sides compared to the standard model. It is not possible to remove the standard carry handle to install a rack-mount kit. The entire side panel must be swapped. | ||
==Specifications== | ==Specifications== | ||
[[Category:Specifications needed]] ''please add'' | [[Category:Specifications needed]] ''please add'' | ||
Input Voltage 115 V or 230 V nominal (selected with an internal jumper) with options for high, medium or low (100 V 110 V 120 V) or (200 V 220 V 240 V) using a second internal jumper. | Input Voltage 115 V or 230 V nominal (selected with an internal jumper) with options for high, medium or low (100 V 110 V 120 V) or (200 V 220 V 240 V) using a second internal jumper. | ||
Fan Flow Rate is higher in units with the 02 (rack-mount) option | Fan Flow Rate is higher in units with the 02 (rack-mount) option | ||
Revision as of 02:43, 1 October 2021
The Tektronix TM5006 is a six-bay mainframe with a GPIB interface for the TM5000 system.
The rightmost compartment (slot 6) can supply high-powered plugins.
The DC rails to the plugins are provided by a switchmode PSU. The supply is not universal input voltage, the input voltage must be set with two internal jumpers. DC rails are 8 V, +26 V, and −26 V. The case has a cooling fan which is connected to the primary of the transformer.
Available options include Opt.02, Rear Inputs, and Opt.10, a set of 19-inch rackmount slides.
For option 02, six BNC connectors and a 50-pin female DD-50 D-sub connector are installed on the rear panel and pin headers soldered internally, but no internal wiring is pre-installed.
The rack-mount kit has different sides compared to the standard model. It is not possible to remove the standard carry handle to install a rack-mount kit. The entire side panel must be swapped.
Specifications
please add
Input Voltage 115 V or 230 V nominal (selected with an internal jumper) with options for high, medium or low (100 V 110 V 120 V) or (200 V 220 V 240 V) using a second internal jumper.
Fan Flow Rate is higher in units with the 02 (rack-mount) option