067-0502-00: Difference between revisions
No edit summary |
No edit summary |
||
Line 18: | Line 18: | ||
{{BeginSpecs}} | {{BeginSpecs}} | ||
{{Spec | Output | {{Spec | Output | DC+, DC-, Squarewave}} | ||
{{Spec | Amplitude | {{Spec | Amplitude range | 0.2 mV to 100 V in 1−2−5 sequence, 5 mA in current loop}} | ||
{{Spec | Amplitude | {{Spec | Amplitude accuracy | Within 0.25% into 1 MΩ load}} | ||
{{Spec | Squarewave | {{Spec | Squarewave repetition rate | 1 kHz within 20% }} | ||
{{Spec | Maximum | {{Spec | Maximum input voltage | 100 V }} | ||
{{Spec | 100x Amplifier | {{Spec | 100x Amplifier gain accuracy | ±1% }} | ||
{{Spec | 100x Amplifier | {{Spec | 100x Amplifier risetime | < 35 μs }} | ||
{{Spec | 100x Amplifier | {{Spec | 100x Amplifier output range | ≥ ±20 V, ±5 mA }} | ||
{{Spec | 100x Amplifier Output | {{Spec | 100x Amplifier Output resistance | approx. 250 Ω}} | ||
{{EndSpecs}} | {{EndSpecs}} | ||
Revision as of 08:23, 21 January 2022
The Tektronix 067-0502-00 is an instrument designed for calibrating oscilloscope calibrators. It is very similar to the 067-0502-01.
Key Specifications
Output | DC+, DC-, Squarewave |
---|---|
Amplitude range | 0.2 mV to 100 V in 1−2−5 sequence, 5 mA in current loop |
Amplitude accuracy | Within 0.25% into 1 MΩ load |
Squarewave repetition rate | 1 kHz within 20% |
Maximum input voltage | 100 V |
100x Amplifier gain accuracy | ±1% |
100x Amplifier risetime | < 35 μs |
100x Amplifier output range | ≥ ±20 V, ±5 mA |
100x Amplifier Output resistance | approx. 250 Ω |
Comparison with 067-0502-01
Sherman Foy contributed the following information:
TekWiki asks the question: What is the difference? I have both units and have done an extensive review.
Looking at the front panels [TNX 4 pix] of the –00 & the –01, it seems that the –01 dropped the × 100 Amplifier section. The front panel is re-laid out, but what remains seems to be identical to the corresponding sections from the –00. In other words, the –00 provides a superset of the functionality of the -01.
Note that the –00 transformer [T310] has 6 windings and the –01 transformer [also T310] has 5 windings. Some currents are different, as indicated.
I have made my own notes for the project of working on the two. Using the schematics for organization, I arrived at the following table. I am using a re-organization of the sheets, the block diagram of the –00 and this table:
Section -00 -01 Sheet in -00 Sheet in -01 Sheet in proposed consolidated drawing set Square Wave calibrator Used in both Used in both 1 1 [s/b 1] × 100 Amplifier Used in -00 Deleted in -01 2 -- [s/b 2] Power Supply -00 has more sections Than –01 3 2 [s/b 3A for -00 & 3B for -01] T310 P/N 120-0442-00 P/N 120-0615-00 3 2 3A/3B High voltage, B+ 285 VRMS, 35 mA 300 VDC, 225 VDC & 100 VDC 125 VRMS, 2 mA 100 VDC only 3 2 3A/3B 6.3 VAC (Chopper) Used in -00 Used in -01, noted as 1.84 A 3 2 3A/3B 100 VRMS (-100 VDC) Used in -00, 40 mA Used in -01, 31 mA 3 2 3A/3B 6.3 VAC (MV 6DJ8) Used in -00 Used in -01 3 2 3A/3B 330 VRMS (Floating Supply) Used in -00 Used in -01 3 2 3A/3B 20 VRMS (× 100 Filaments) Used in -00 Not used in –01 T310 3 2 3A/3B Floating power supply Used in both; drives Square Wave MV Used in both; drives Square Wave MV 4 3 [s/b 4]