2710: Difference between revisions
No edit summary |
No edit summary |
||
Line 14: | Line 14: | ||
* [[Media:Tek 2710 option 04 manual change info.pdf|Tektronix 2710 Option 04 Manual Change Information]] | * [[Media:Tek 2710 option 04 manual change info.pdf|Tektronix 2710 Option 04 Manual Change Information]] | ||
* [[Media:Mod 2710F04 opt04 internal tracking gen.pdf]] | * [[Media:Mod 2710F04 opt04 internal tracking gen.pdf]] | ||
* [[Media:Tek 2710 normalizations.pdf|Tektronix 2710 Normalizations]] | |||
}} | }} | ||
The '''Tektronix 2710''' is a spectrum analyzer with a frequency range of 10 kHz to 1.8 GHz. | The '''Tektronix 2710''' is a spectrum analyzer with a frequency range of 10 kHz to 1.8 GHz. |
Revision as of 14:33, 9 December 2022
The Tektronix 2710 is a spectrum analyzer with a frequency range of 10 kHz to 1.8 GHz.
Key Specifications
Frequency | 10 kHz to 1.8 GHz |
---|---|
Frequency Span | 10 kHz/div to 100 MHz/div in 1–2–5 sequence, plus 0 Hz and MAX: 180 MHz/div |
Resolution Bandwidth | 3 kHz, 30 kHz, 300 kHz, and 5 MHz (–6 dB); 300 Hz with Opt. 01 |
RF Input | 50 Ω, max. +20 dBm (0.1 W or 2.2 V) CW, ≤100 V DC (blocking capacitor) |
RF Attenuator | 0 dB to 50 dB, 2 dB steps |
Sensitivity | –117 dBm at 3 kHz RBW, –129 dBm at 3 kHz RBW w/preamp |
Reference Level | –70 dBm to +20 dBm; –23 dBmV to +66.9 dBmV |
Sweep Speed | 2 sec/Div to 1 µs/Div in 1–2–5 sequence |
Triggering Modes | Free Run, Internal, External, Line, TV Line, and TV Field |
Displayed Average Noise | −66 dBm to −100 dBm |
Display Dynamic Range | 80 dB |
Dynamic Range | ≥100 dBm |
Internal Calibrator | 100 MHz marker at –30 dBm |
Weight | Standard accessories: 11.25 kg (<25 lbs); Basic configuration: 9.5 kg (<21 lbs) |
Power | 90 − 132 VAC, 48 to 440 Hz; 90 – 250 VAC, 47 to 63 Hz, 90 W (standard) / 105 W max (with options) |
Operating Environment | 0 °C to +50 °C; Humidity: MIL-T-28800E; Altitude: Up to 4.6 km (15,000 ft) |
Features |
|
Options
- Opt. 01: 300 Hz Resolution Bandwidth filter; Phase-lock with span/div ≤20 kHz
- Opt. 02: Frequency Counter, 1 Hz or 1 kHz, menu selectable
- Opt. 03: GPIB Interface (cannot be combined with Opt. 09)
- Opt. 04: TG Tracking Generator, 100 kHz to 1.8 GHz (cannot be combined with Opt. 14)
- Opt. 06: Battery Power Option (2704 Inverter and 2705 Battery Pack)
- Opt. 08: Replace GPIB with RS-232-C (cannot be combined with Opt. 03)
- Opt. 09: Centronix interface (cannot be combined with Opt. 03)
- Opt. 10: Video Monitor mode
- Opt. 11: Non-volatile RAM storage for setups, displays, and user-definable key sequences
- Opt. 14: Adds Resolution Bandwidths of 1 kHz, 10 kHz, 100 kHz, and 1 MHz (cannot be combined with Opt. 04)
- Opt. 15: Tek 1405 TV Sideband Adapter interface installed at J101 as SMA female
- Opt. 30: Rackmount
- Opt. 33: Travel Line Package
- Opt. 34: Portable to Rackmount Adapter.
See also the 2706 preselector and 2707 tracking generator.
Connectors
Front panel: RF INPUT N connector (50Ω, +20 dBm Max, 100 V DC blocking capacitor); and GEN N connector (50Ω, 0 to -48 dBm, Opt. 4 Tracking Generator generates a sweep frequency to enable the Spectrum Analyzer to make circuit response measurements, i.e. scalar network analysis).
Rear panel: J101 SMA connector (Female SMA installed for Opt. 15), J102 EXT TRIG BNC connector (10 kΩ, 35V (DC + peak AC)), J103 DB9 connector (Not RS-232 compatible. Accessory connector: Pin 1 Video Input, Pin 2 Ground, Pin 3 Video Output, Pin 6 Sweep Gate, Pin 7 Sweep Output, all others unused), J104 GPIB connector
Originally included accessories
Operator’s Manual; Power Cord; Front Cover; Adapter 50 Ω N Male to BNC Female, 75 Ω to 50 Ω Minimum Loss Attenuator N Male to BNC Female