PG506: Difference between revisions
Jump to navigation
Jump to search
No edit summary |
No edit summary |
||
Line 5: | Line 5: | ||
* [[Media:070-3383-00.pdf|Tektronix PG506 Instruction Manual]] SN 40000+ (whole schematics, OCR) | * [[Media:070-3383-00.pdf|Tektronix PG506 Instruction Manual]] SN 40000+ (whole schematics, OCR) | ||
'''PG506A''' | '''PG506A''' | ||
* [[Media:070-6687-00.pdf|Tektronix PG506A Instruction Manual | * [[Media:070-6687-00.pdf|Tektronix PG506A Instruction Manual]] | ||
}} | }} | ||
It functionally replaces the [[067-0502-01]] in the "standard amplitude" mode, and the [[106]] in the "fast rise" and "high amplitude" modes. | It functionally replaces the [[067-0502-01]] in the "standard amplitude" mode, and the [[106]] in the "fast rise" and "high amplitude" modes. |
Revision as of 04:22, 15 October 2022
The Tektronix PG506 is a 1 MHz calibration generator plug-in for the TM500 system.
It functionally replaces the 067-0502-01 in the "standard amplitude" mode, and the 106 in the "fast rise" and "high amplitude" modes.
There are (at least) two versions of the PG506. The older version has two relays that switch a low-pass (noise) filter into the 10 V and below ranges. The new version just has a capacitor across the output for these ranges.
Common failure modes: In case of low or no output on the 20 V and above ranges, check Q270, Q280 and Q290.
Links
Pictures
PG506 older version
PG506 newer version
-
-
PG506 Left Internal
-
PG506 Left Internal
-
PG506 Right Internal
-
PG506 Right Internal
PG506A
-
PG506A Front
-
Newer version without filter relays
-
PG506A
-
PG506A output waveform