Neon Lamps: Difference between revisions

From TekWiki
Jump to navigation Jump to search
No edit summary
No edit summary
Line 23: Line 23:
==See also==
==See also==
* [[wikipedia:Neon_lamp|Neon lamp]] @ Wikipedia
* [[wikipedia:Neon_lamp|Neon lamp]] @ Wikipedia
* [[Media:GE-Glow-lamp-manual-Theory-Circuits-Ratings-2nd-Edition-1965-118-pages.pdf|GE Glow Lamp Manual, 2nd Ed. 1965]]
* William G. Miller, ''[[Media:Miller, Using and Understanding Miniature Neon Lamps (1969).pdf|Using and Understanding Miniature Neon Lamps]]'' (1969)
* William G. Miller, ''[[Media:Miller, Using and Understanding Miniature Neon Lamps (1969).pdf|Using and Understanding Miniature Neon Lamps]]'' (1969)
* [[:Category:Nixie tubes|Nixie tubes]]
* [[:Category:Nixie tubes|Nixie tubes]]
Line 46: Line 47:
|align="center"| 58 V
|align="center"| 58 V
|align="center"| 0.3 mA
|align="center"| 0.3 mA
| =~ modern type [https://mouser.com/ProductDetail/VCC/A1A?qs=rHu0DOexXLD2XDJrE%252BAsOw%3D%3D A1A]
|-
|-
| NE-2V
| NE-2V
Line 52: Line 54:
|align="center"| 58 (46-78) V
|align="center"| 58 (46-78) V
|align="center"| 0.6 mA
|align="center"| 0.6 mA
| short length, formed tip for end-on viewing, mild radioactive additive for the reduction of dark effect
| short length, formed tip for end-on viewing, mild radioactive additive for the reduction of dark effect;
| =~ modern type [https://eu.mouser.com/ProductDetail/VCC/A2B?qs=%252BE2FUV3Ulmz0JmjQi2dc2Q%3D%3D A2B]
|-
|-
| NE-23
| NE-23

Revision as of 03:21, 6 June 2023

Neon lamps are used for at least three different things in Tek instruments:

  1. Indicator lamps, e.g., for power, beam position, uncalibrated status, or a decimal point
  2. Voltage clamp, similar to how a zener diode is used
  3. In relaxation oscillators, e.g. in 7000-series power supplies
Broken lead (left) on neon in 585 time base

Neon lamps fail more than most other components. A common failure mode is for the leads to break off, often close to the glass bulb, due to corrosion and vibration. Another failure mode is for the I-V characteristic to drift.

Neon lamps are used as a voltage clamp in the CRT circuit for setting the CRT grid voltage in many Tek instruments. A typical example is seen in the 561A:

If the neon lamps fail, the CRT grid voltage will be affected.

Another common use is for DC bias shifting, e.g. in 500-series timebase circuits.


When in doubt, replace the neon lamps. In a pinch, it may be possible to reattach broken leads after careful grinding of the glass.

See also

Types and data

Type Tek P/N Voltage Current Notes
Ignition Operating Nominal
NE-2 150-002 76 (60−90) V 58 V 0.3 mA =~ modern type A1A
NE-2V 150-030 76 (60−90) V 58 (46-78) V 0.6 mA short length, formed tip for end-on viewing, mild radioactive additive for the reduction of dark effect; =~ modern type A2B
NE-23 150-027 74 (60−90) V 59 V 0.3 mA mild radioactive additive for reduction of dark effect

Please add