2710: Difference between revisions

From TekWiki
Jump to navigation Jump to search
No edit summary
No edit summary
Line 67: Line 67:


===Connectors===
===Connectors===
Front panel: RF INPUT [[N connector]] (50Ω, +20 dBm Max, 100 V DC blocking capacitor); and GEN [[N connector]] (50Ω, 0 to -48 dBm, Opt. 4 Tracking Generator generates a sweep frequency to enable the Spectrum Analyzer to make circuit response measurements, i.e. scalar network analysis).
'''Front panel:''' RF INPUT [[N connector]] (50 Ω, +20 dBm Max, 100 V DC blocking capacitor); GEN [[N connector]] (50 Ω, 0 to -48 dBm, Opt. 4 Tracking Generator generates a sweep frequency to enable the Spectrum Analyzer to make circuit response measurements, i.e. scalar network analysis).


Rear panel: J101 [[SMA connector]] (Female SMA installed for Opt. 15), J102 EXT TRIG [[BNC connector]] (10 kΩ, 35V (DC + peak AC)), J103 [[DB9 connector]] (Not RS-232 compatible.  Accessory connector: Pin 1 Video Input, Pin 2 Ground, Pin 3 Video Output, Pin 6 Sweep Gate, Pin 7 Sweep Output, all others unused), J104 [[GPIB connector]]
'''Rear panel:''' J101 [[SMA connector]] (Female SMA installed for Opt. 15), J102 EXT TRIG [[BNC connector]] (10 kΩ, 35 V (DC + peak AC)), J103 [[DB9 connector]] (Accessory connector, not RS-232: Pin 1 Video Input, Pin 2 Ground, Pin 3 Video Output, Pin 6 Sweep Gate, Pin 7 Sweep Output, all others unused), J104 [[GPIB connector]]


===Originally included accessories===
===Originally included accessories===

Revision as of 08:33, 12 October 2023

Tektronix 2710
1.8 GHz Spectrum Analyzer
Tektronix 2710

Produced from 1987 to 1993

Manuals
Manuals – Specifications – Links – Pictures

The Tektronix 2710 is a spectrum analyzer with a frequency range of 10 kHz to 1.8 GHz.

Key Specifications

Frequency 10 kHz to 1.8 GHz
Frequency Span 10 kHz/div to 100 MHz/div in 1–2–5 sequence, plus 0 Hz and MAX: 180 MHz/div
Resolution Bandwidth 3 kHz, 30 kHz, 300 kHz, and 5 MHz (–6 dB); 300 Hz with Opt. 01
RF Input 50 Ω, max. +20 dBm (0.1 W or 2.2 V) CW, ≤100 V DC (blocking capacitor)
RF Attenuator 0 dB to 50 dB, 2 dB steps
Sensitivity –117 dBm at 3 kHz RBW, –129 dBm at 3 kHz RBW w/preamp
Reference Level –70 dBm to +20 dBm; –23 dBmV to +66.9 dBmV
Sweep Speed 2 sec/Div to 1 µs/Div in 1–2–5 sequence
Triggering Modes Free Run, Internal, External, Line, TV Line, and TV Field
Displayed Average Noise −66 dBm to −100 dBm
Display Dynamic Range 80 dB
Dynamic Range ≥100 dBm
Internal Calibrator 100 MHz marker at –30 dBm
Weight Standard accessories: 11.25 kg (<25 lbs); Basic configuration: 9.5 kg (<21 lbs)
Power 90 − 132 VAC, 48 to 440 Hz; 90 – 250 VAC, 47 to 63 Hz, 90 W (standard) / 105 W max (with options)
Operating Environment 0 °C to +50 °C; Humidity: MIL-T-28800E; Altitude: Up to 4.6 km (15,000 ft)
Features
  • 28KB non-volatile memory (NVRAM) for up to 18 displays of waveforms and markers, and up to 9 front panel setups, user-defined key programs, and antenna factor tables
  • Single and delta markers for readout of frequency and amplitude
  • analog display capability, TV Line and TV Field triggering to provide demodulation of video carriers
  • internal audio amplifier and AM/FM detectors
  • digital storage (positive peak only, positive/negative peak, max hold, waterfall)
  • alternate reference level units (dBm, dBmV, dBV, dBµV, dBµW, and dBµV/m)
  • direct plot/print via Opt. 03 GPIB or Centronics interface Opt. 09

Options

  • Opt. 01: 300 Hz Resolution Bandwidth filter; Phase-lock with span/div ≤20 kHz
  • Opt. 02: Frequency Counter, 1 Hz or 1 kHz, menu selectable
  • Opt. 03: GPIB Interface (cannot be combined with Opt. 09)
  • Opt. 04: TG Tracking Generator, 100 kHz to 1.8 GHz (cannot be combined with Opt. 14)
  • Opt. 06: Battery Power Option (2704 Inverter and 2705 Battery Pack)
  • Opt. 08: Replace GPIB with RS-232-C (cannot be combined with Opt. 03)
  • Opt. 09: Centronix interface (cannot be combined with Opt. 03)
  • Opt. 10: Video Monitor mode
  • Opt. 11: Non-volatile RAM storage for setups, displays, and user-definable key sequences
  • Opt. 14: Adds Resolution Bandwidths of 1 kHz, 10 kHz, 100 kHz, and 1 MHz (cannot be combined with Opt. 04)
  • Opt. 15: Tek 1405 TV Sideband Adapter interface installed at J101 as SMA female
  • Opt. 30: Rackmount
  • Opt. 33: Travel Line Package
  • Opt. 34: Portable to Rackmount Adapter.

See also the 2706 preselector and 2707 tracking generator.

Connectors

Front panel: RF INPUT N connector (50 Ω, +20 dBm Max, 100 V DC blocking capacitor); GEN N connector (50 Ω, 0 to -48 dBm, Opt. 4 Tracking Generator generates a sweep frequency to enable the Spectrum Analyzer to make circuit response measurements, i.e. scalar network analysis).

Rear panel: J101 SMA connector (Female SMA installed for Opt. 15), J102 EXT TRIG BNC connector (10 kΩ, 35 V (DC + peak AC)), J103 DB9 connector (Accessory connector, not RS-232: Pin 1 Video Input, Pin 2 Ground, Pin 3 Video Output, Pin 6 Sweep Gate, Pin 7 Sweep Output, all others unused), J104 GPIB connector

Originally included accessories

Operator’s Manual; Power Cord; Front Cover; Adapter 50 Ω N Male to BNC Female, 75 Ω to 50 Ω Minimum Loss Attenuator N Male to BNC Female

Links

Documents Referencing 2710

Document Class Title Authors Year Links
26W-7045.pdf Application Note Random Noise Measurement with the Spectrum Analyzer Morris Engelson 1989

Pictures