84: Difference between revisions

From TekWiki
Jump to navigation Jump to search
No edit summary
No edit summary
Line 33: Line 33:
==Pictures==
==Pictures==
<gallery>
<gallery>
Tel_84_Front.jpeg | Front
Tek 067-0523-00 front.jpg  | 067-0523-00 Front
067-0523-00 cat.jpg
Tel_84_Front.jpeg         | Type 84 Front
Tek 067-0523-00.jpg
Tek 067-0523-00 1.jpg | 067-0523-00
Tek 067-0523-00 2.jpg | 067-0523-00
Tek 067-0523-00 3.jpg | 067-0523-00
Tek 067-0523-00 4.jpg | 067-0523-00
Tek 84 1.jpg
Tek 84 2.jpg
Tek 84 3.jpg
Tek 84 4.jpg
Tek 067-0523-00 internal1.jpg | 067-0523-00
Tek 067-0523-00 internal2.jpg | 067-0523-00
Tek 067-0523-00 rear.jpg      | 067-0523-00
</gallery>
 
'''Schematics'''
<gallery>
Type-84-schem.png
Type-84-schem.png
Tek 067-0523-00.jpg
Tek 067-0523-00 1.jpg|067-0523-00
Tek 067-0523-00 2.jpg|067-0523-00
Tek 067-0523-00 3.jpg|067-0523-00
Tek 067-0523-00 4.jpg|067-0523-00
Tek 84 1.jpg|84
Tek 84 2.jpg|84
Tek 84 3.jpg|84
Tek 84 4.jpg|84
Tek 067-0523-00 internal1.jpg|067-0523-00
Tek 067-0523-00 internal2.jpg|067-0523-00
Tek 067-0523-00 rear.jpg|067-0523-00
</gallery>
</gallery>



Revision as of 09:09, 19 November 2023

Tektronix 84
calibration fixture
Type 84

Compatible with 580-series scopes

Produced from 1962 to (?)

Manuals
Manuals – Specifications – Links – Pictures

The Tektronix Type 84 is a calibration fixture plug-in for 580-series scopes.

It was renamed the 067-0523-00 during production (SN 974+).

The Type 84 provides a calibrated amplitude signal for adjusting gain, a fast-rise voltage step – produced by a mercury switch – for adjusting the impulse response of the vertical amplifiers in the scope, and load resistors to test the power supply of the scope. It also features an external input to connect a time-mark generator during calibration.

Key Specifications

Display signal Calibration signal, fast-rise pulses, or external
Calibration reference signal 200 mV, front-panel check
Pulse amplitude Adjustable to plus or minus 4 cm deflection
Pulse repetition rate 550 to 750 pulses per second