7000-series remote control connector: Difference between revisions
Jump to navigation
Jump to search
No edit summary |
(7503 is similar) |
||
Line 1: | Line 1: | ||
The '''remote control connector''' on the rear of the [[7904]] | The '''remote control connector''' on the rear of the [[7904]] and [[7503]] scopes is an [[Amphenol 165 series connectors|Amphenol 165-16]] socket. | ||
{| class="wikitable" | {| class="wikitable" | ||
Line 13: | Line 13: | ||
| C || A Remote ready || +5 V via 47 Ω when active, can drive #49 bulb | | C || A Remote ready || +5 V via 47 Ω when active, can drive #49 bulb | ||
|- | |- | ||
| D || B Remote ready || +5 V via 47 Ω when active, can drive #49 bulb | | D || B Remote ready (*) || +5 V via 47 Ω when active, can drive #49 bulb | ||
|- | |- | ||
| E || Remote readout lock-out || Close to ground within 0.1 V, 2 mA max, 2 V open voltage | | E || Remote readout lock-out (*) || Close to ground within 0.1 V, 2 mA max, 2 V open voltage | ||
|- | |- | ||
| F || Remote single-shot lock-out || Closure to ground within 0.4 V, pin E was grounded before, rate of change ≥ 0.1 V/μs, 3 mA max, 10 V open voltage | | F || Remote single-shot lock-out (*) || Closure to ground within 0.4 V, pin E was grounded before, rate of change ≥ 0.1 V/μs, 3 mA max, 10 V open voltage | ||
|- | |- | ||
| H || n/c || | | H || n/c || | ||
Line 27: | Line 27: | ||
|} | |} | ||
:'' (*) 7904 only'' | |||
[[Category:Instrument specific interfaces]] | [[Category:Instrument specific interfaces]] |
Revision as of 05:07, 24 January 2021
The remote control connector on the rear of the 7904 and 7503 scopes is an Amphenol 165-16 socket.
Pin | Function | Notes |
---|---|---|
A | Single sweep reset | Close to ground within 0.5 V, 10 mA max, minimum 10 μs pulse width |
B | Ground | |
C | A Remote ready | +5 V via 47 Ω when active, can drive #49 bulb |
D | B Remote ready (*) | +5 V via 47 Ω when active, can drive #49 bulb |
E | Remote readout lock-out (*) | Close to ground within 0.1 V, 2 mA max, 2 V open voltage |
F | Remote single-shot lock-out (*) | Closure to ground within 0.4 V, pin E was grounded before, rate of change ≥ 0.1 V/μs, 3 mA max, 10 V open voltage |
H | n/c | |
J | n/c | |
K | n/c |
- (*) 7904 only