P6205: Difference between revisions
m (Text replacement - "Category:Active Oscilloscope Probes" to "Category:Active oscilloscope probes") |
(manual links changed to internal docs) |
||
Line 15: | Line 15: | ||
==Manual== | ==Manual== | ||
* [ | * [[Media:070-8202-01.pdf|Tektronix P6205 Manual (PDF)]] | ||
* [http://w140.com/tekp6203-p62204-p62205-p6217-p6231.pdf Tektronix P6203 P6204 P6205 P6217 P6231 Catalog Specs (PDF)] | * [http://w140.com/tekp6203-p62204-p62205-p6217-p6231.pdf Tektronix P6203 P6204 P6205 P6217 P6231 Catalog Specs (PDF)] | ||
Revision as of 11:59, 23 May 2021
The Tektronix P6205 is a 750 MHz active FET probe for TDS500 series, DSA600 series, CSA404 and 11000-series scopes with Tekprobe BNC inputs.
Key Specifications
Bandwidth | 750 MHz (probe only) 600 MHz on 11A72 450 MHz on TDS500/600 |
---|---|
Rise time | < 467 ps |
Attenuation | 10× |
Input impedance | 1 MΩ // 2 pF |
Output | into 50 Ω |
Dynamic range | ±10 V |
Max. input voltage | 40 V DC + peak AC non-destructive |
Interface | Tekprobe BNC connector |
Manual
Pictures
-
Exploded view of the probe assembly. The PCB is soldered to the shielding tube and the coax, and the tube is then crimped over the coax.
-
The probe uses "lossless" 50 Ω coax, as there's amplification at the business end. The circuitry connects to the TekProbe interface to declare the probe's identity to the scope.
-
The tip network comprises three laser-trimmed resistors and a laser-trimmed capacitor. The tip resistance measures 800 kΩ, the middle resistor was on the order of 25 Ω, and the resistance to ground is 200 kΩ. This makes a 1/5 divider from tip to the amplifier.
-
-
-
It looks like the amplifier has two identical sections, presumably the same chip is used for differential active probes.