7B15: Difference between revisions
No edit summary |
No edit summary |
||
Line 8: | Line 8: | ||
|discontinued=1990 | |discontinued=1990 | ||
|series=7000-series scopes | |series=7000-series scopes | ||
|designers=Art Metz;Bruce Hofer | |designers=Art Metz;Bruce Hofer;Val Garuts | ||
|manuals= | |manuals= | ||
* [http://bama.edebris.com/download/tek/7b15/tek-7b15.pdf Tektronix 7B15 Manual (PDF)] | * [http://bama.edebris.com/download/tek/7b15/tek-7b15.pdf Tektronix 7B15 Manual (PDF)] |
Revision as of 04:45, 15 August 2021
The Tektronix 7B15 is a 1 GHz delaying time base for 7000-series scopes, in particular, the 1 GHz 7104 (typically used with 7A29 vertical amplifiers and a 7B10 delayed time base).
The 7B15 provides digital readout of delay and delta times. In effect, the 7B15/7B10 pair are speed-improved versions of the 7B85/7B80.
Key Specifications
Sweep | 200 ms/Div to 2 ns/Div in 1−2−5 sequence, ×10 magnifier down to 200 ps/Div, variable to ×2.5 |
---|---|
Triggering | to 1 GHz |
Jitter | less than 30 ps at 1 GHz |
Features |
|
Internals
A ramp generator is started when the trigger signal arrives. The delayed sweep is started when a comparator detects that the ramp has reached a control voltage, set by the front panel knobs. The control voltage is therefore proportional to the amount of delay and is measured using a 155-0185-00 4-digit DVM circuit, then displayed on the oscilloscope readout. The display, scaled in seconds, provides the operator with a precise readout of the delay.
Other custom Tek ICs used: 155-0150-00 trigger detector, 155-0160-00 trigger amplifier.
The 7B15 uses the +50 V supply provided by the mainframe.
Pictures
-
7B15 front panel
-
left side
-
right side
-
Family resemblance: 7B15 (top) vs. 7B85 (bottom). Left side.
-
Family resemblance: 7B15 (top) vs. 7B85 (bottom). Right side.
-
-