{{Spec | Input impedance  | 10 MΩ // 2 pF with or without additional 10× attenuator tip, +2 pF for AC coupling tip }}
{{Spec | Input impedance  | 10 MΩ // 2 pF with or without additional 10× attenuator tip, +2 pF for AC coupling tip }}
{{Spec | Input range      | ± 6 V (± 60 V with additional 10× attenuator tip) }}
{{Spec | Input range      | ± 6 V (± 60 V with additional 10× attenuator tip) }}
{{Spec | DC Offset range  | ± 55 V (± 200 V with additional 10× attenuator tip) }}
{{Spec | DC offset        | ± 55 V (± 200 V with additional 10× attenuator tip) }}
{{Spec | Max input        | ± 200 V (nondestructive) }}
{{Spec | Max input        | ± 200 V (nondestructive) }}
{{Spec | Cable length    | 2 m (6.5') }}
{{Spec | Cable length    | 2 m (6.5') }}

Revision as of 09:06, 21 January 2022

Tektronix P6202
500 MHz Active FET Probe
Tektronix P6202

Compatible with 50 Ω scope inputs

Produced from 1976 to (?)

Manuals


Manuals – Specifications – Links – Pictures

The Tektronix P6202 is a 500 MHz active 10× FET probe with 50 Ω output impedance, introduced in 1976.

The 6202 has its own mains power supply, the 6202A (1980) has a LEMO S-series connector fitting the 7000-series scopes' probe power outlets.

Key Specifications

Bandwidth DC to 500 MHz (16 Hz lower limit with additional AC coupling tip)
Rise time < 0.7 ns
Attenuation 10 : 1 (100 : 1 with additional 10× attenuator tip, 010-0384-00)
Input impedance 10 MΩ // 2 pF with or without additional 10× attenuator tip, +2 pF for AC coupling tip
Input range ± 6 V (± 60 V with additional 10× attenuator tip)
DC offset ± 55 V (± 200 V with additional 10× attenuator tip)
Max input ± 200 V (nondestructive)
Cable length 2 m (6.5')

Internals

An initial 8 MΩ : 2 MΩ divider drives a dual JFET (source follower + current source), a PNP emitter follower and finally an NPN emitter follower as a cable driver. These stages use SMD components located in the probe tip. The probe control box at the end of the coax only contains a ±7 V regulator, a switchable 50 Ω terminator, and the offset potentiometers that supply ±15 V to the bottom of the input divider. There is no input protection apart from the divider.

Pictures

P6202

P6202A

Measurements

Rise time of a P6202 500 MHz FET probe measured on a 7S12 plugin in a 7844 mainframe (with 25 ps S-52 pulse generator head and 25 ps S-4 sampling head). Measured rise time ~440 ps observed on this specimen is equivalent to 795 MHz bandwidth.