410: Difference between revisions

From TekWiki
Jump to navigation Jump to search
(Bob Shand did the mechanical design (see Tekscope May 1972))
No edit summary
Line 11: Line 11:
|designers=Bob Shand;
|designers=Bob Shand;
|manuals=
|manuals=
* ''Manual needed - please add'' [[Category:Manual needed]]
* [https://archive.org/download/070-0658-01/070-0658-01.pdf Tektronix 410 Manual]
* [[Media:Tek 410 schematics.pdf|Tektronix 410 Schematics (PDF)]]
* [[Media:Tek 410 schematics.pdf|Tektronix 410 Schematics (PDF)]]
* Factory Calibration Procedures: [[Media:Tek 016-0107-00 fcp aug 1967.pdf|August 1967]] / [[Media:Tek 410 fcp oct 1967.pdf|October 1967]] / [[Media:Tek 015-0102-00 fcp may 1968.pdf|May 1968]]  
* Factory Calibration Procedures: [[Media:Tek 016-0107-00 fcp aug 1967.pdf|August 1967]] / [[Media:Tek 410 fcp oct 1967.pdf|October 1967]] / [[Media:Tek 015-0102-00 fcp may 1968.pdf|May 1968]]  

Revision as of 03:12, 22 August 2022

Tektronix 410
Physiological monitor
Tek 410 Physiological Monitor

Produced from 1967 to (?)

Manuals
(All manuals in PDF format unless noted otherwise)
Manuals – Specifications – Links – Pictures

The Tektronix 410 Physiological Monitor is a specialized oscilloscope for medical applications.

It is a small, battery-powered instrument using a magnetically deflected CRT, providing ECG, EEG and pulse measurement modes.

The graticule includes a heart rate scale allowing readings in beats/min (at 50 mm/s).

For portable operation, a 016-0107-02 battery pack (10×NiCd, charger) was available.

The 067-0549-00 Input Adapter can be used to calibrate the 410.

Key Specifications

Sweep 25, 50 or 100 mm/s
Bandwidth 0.1 to 250 Hz (ECG, AUX) or 100 Hz (EEG)
Deflection 10 mm/50 μV (EEG), 20 mm/mV (ECG), 2 mm/mV (AUX)
Input impedance 2 MΩ (EEG, ECG) / 20 MΩ (AUX)
CMRR 150,000 : 1
Battery 10 size C NiCd cells, operating time 8-12 hours, charging 14-16 hours
CRT 5" magnetically deflected, P-7 phosphor

Links

Photos