PG506: Difference between revisions
Jump to navigation
Jump to search
No edit summary |
No edit summary |
||
Line 15: | Line 15: | ||
==Links== | ==Links== | ||
* [https://www.youtube.com/watch?v=IschnXSgOqc Zenwizard Studios - PG506 Repair and Calibration] | * [https://www.youtube.com/watch?v=IschnXSgOqc Zenwizard Studios - PG506 Repair and Calibration] | ||
* [https://www.youtube.com/watch?v=-Q74hj8hA5g Zenwizard Studios - Tektronix PG506A Refresh and Checkin] | |||
==Pictures== | ==Pictures== |
Revision as of 08:04, 31 August 2023
The Tektronix PG506 is a 1 MHz calibration generator plug-in for the TM500 system.
It functionally replaces the 067-0502-01 in the "standard amplitude" mode, and the 106 in the "fast rise" and "high amplitude" modes.
There are (at least) two versions of the PG506. The older version has two relays that switch a low-pass (noise) filter into the 10 V and below ranges. The new version just has a capacitor across the output for these ranges.
Common failure modes: In case of low or no output on the 20 V and above ranges, check Q270, Q280 and Q290.
Links
- Zenwizard Studios - PG506 Repair and Calibration
- Zenwizard Studios - Tektronix PG506A Refresh and Checkin
Pictures
PG506 older version
PG506 newer version
-
-
PG506 Left Internal
-
PG506 Left Internal
-
PG506 Right Internal
-
PG506 Right Internal
PG506A
-
PG506A Front
-
Newer version without filter relays
-
PG506A
-
PG506A output waveform