PG506

From TekWiki
Revision as of 05:00, 1 October 2021 by Peter (talk | contribs)
Jump to navigation Jump to search
Tektronix PG506
1 MHz calibration generator
Tektronix PG506

Produced from 1974 to 1993

Manuals
Manuals – Specifications – Links – Pictures

The Tektronix PG506 is a 1 MHz calibration generator plug-in for the TM500 system.

It functionally replaces the 067-0502-01 in the "standard amplitude" mode, and the 106 in the "fast rise" and "high amplitude" modes.

There are (at least) two versions of the PG506. The older version has two relays that switch a low-pass (noise) filter into the 10 V and below ranges. The new version just has a capacitor across the output for these ranges.

Common failure modes: In case of low or no output on the 20 V and above ranges, check Q270, Q280 and Q290.

Pictures

PG506 older version

PG506 newer version

PG506A