Line 5: | Line 5: | ||
Some of these relays have symmetrical socket layouts so that they can be installed in both possible 180° rotated positions. | Some of these relays have symmetrical socket layouts so that they can be installed in both possible 180° rotated positions. | ||
Possible repair strategies include "micro-surgery" (Ref.1) or installing modern miniature relays on top of the cut-off base. | Possible repair strategies include "micro-surgery" (Ref.1) or installing modern miniature gold-contact relays on top of the cut-off base. | ||
{| class="wikitable" | {| class="wikitable" | ||
Line 15: | Line 15: | ||
! Replacement | ! Replacement | ||
|- | |- | ||
| 148-0034-00 || DPDT || 15 V 600 Ω || [[7A11]]: K401,K403,K405,K407,K411,K413,K415,K417,K431<br />[[7A13]]: K47,K48,K480,K490<br />[[7A14]]: K50,K62,K330 || AXICOM V23079-D2003-B301 (Ref. 2)<br />Omron G6N, G6S | | 148-0034-00 || DPDT || 15 V 600 Ω || [[7A11]]: K401,K403,K405,K407,K411,K413,K415,K417,K431<br />[[7A13]]: K47,K48,K480,K490<br />[[7A14]]: K50,K62,K330 || AXICOM V23079-D2003-B301 (Ref. 2)<br />[[Media:Omron G6 relay.jpg|Omron G6N, G6S]] | ||
|- | |- | ||
| 148-0034-03 || DPDT || 15 V 600 Ω || [[7A11]] B100000+: K401,K403,K405,K407,K411,K413,K415,K417,K431 || AXICOM V23079-D2003-B301 (Ref. 2) | | 148-0034-03 || DPDT || 15 V 600 Ω || [[7A11]] B100000+: K401,K403,K405,K407,K411,K413,K415,K417,K431 || AXICOM V23079-D2003-B301 (Ref. 2) | ||
Line 71: | Line 71: | ||
Tek 148-0055-00 4.jpg|148-0055-00 | Tek 148-0055-00 4.jpg|148-0055-00 | ||
tek-148-0107-4.jpg | 148-0107-04 TO-5 replacement relay | tek-148-0107-4.jpg | 148-0107-04 TO-5 replacement relay | ||
Omron G6 relay.jpg | replacement Omron G6 relay soldered onto Tek socket | |||
7a11-probe-interior-back.jpg | Two 148-0058-00 in the [[7A11]] probe | 7a11-probe-interior-back.jpg | Two 148-0058-00 in the [[7A11]] probe | ||
</gallery> | </gallery> |