7A14: Difference between revisions

16 bytes added ,  23 August 2023
no edit summary
No edit summary
No edit summary
Line 27: Line 27:
The P6021 shorts this ring to ground, directly switching three relays connected to −15 V internally. When connecting a [[P6019]] probe, the ring must also be grounded but the probe connector does not have the matching pin.  The [[131-0750-00]] adapter solves this problem.
The P6021 shorts this ring to ground, directly switching three relays connected to −15 V internally. When connecting a [[P6019]] probe, the ring must also be grounded but the probe connector does not have the matching pin.  The [[131-0750-00]] adapter solves this problem.


The front panel gain and low frequency compensation screwdriver adjustments affects gain for either P6021 or P6022, i.e. swapping probes requires re-adjustment. Note that the low frequency compensation adjustment has a large impact on waveform accuracy and requires careful adjustment to a certain slope when using a 1 kHz calibrator signal as discussed in the manual.  
The front panel gain and low frequency compensation screwdriver adjustments do not differentiate between the P6021 or P6022 probes, i.e. swapping probes requires re-adjustment. Note that the low frequency compensation adjustment has a large impact on waveform accuracy and requires careful adjustment to a certain slope when using a 1 kHz calibrator signal as discussed in the manual.  