067-0587-10
The Tektronix 067-0587-10 is a Calibration fixture / signal standardizer for 7000-series scopes. It has the same performance specifications as the 067-0587-02 but is modified for compatibility with the 7612D.
In the 067-0587-10, the signal coming from the PRETRIG OUT generates one trigger for each staircase signal when used with the TEST switch in the GAIN position. The 067-0587-02 generates a pre-trigger signal on the PRETRIG OUT connector while in any mode.
Key Specifications
— GAIN mode — | |
Staircase accuracy | ± 0.3% in ± 3 Div, ± 0.5% in ± 4 Div, Linearity ± 0.2% of 10 Div |
— STEP RESP mode — | |
Amplitude | 2 to 10 Div |
Position | Top at least ±5 Div from center of screen |
Rise time | < 150 ps |
Repetition rate | 10 Hz to 1 MHz in decade steps, 0.1 % |
Pretrig Out | 0.5 V into 50 Ω, 55 to 95 ns lead time (internally variable) |
— FREQ RESP mode — | |
Input impedance | 50 Ω |
Input range | 0.5 to 1 Vp-p |
Flatness | ± 3% from 3 MHz to 1 GHz |
Pictures
-
-
067-0587-10 left side (cam switch assembly on bottom left)
-
067-0587-10 right side − note extra switch and resistor in top middle area of PCB that differentiates it from 067-0587-02
-
067-0587-10 with modified rear plastic interface to fit 11000-series scopes