DM501/Repairs: Difference between revisions

From TekWiki
Jump to navigation Jump to search
No edit summary
No edit summary
Line 3: Line 3:
The contact bar in switches can break. Replacing the contact bar and contacts can be done without desoldering the switches.
The contact bar in switches can break. Replacing the contact bar and contacts can be done without desoldering the switches.
Refer to the the pictures and the [[260-1971-00/Repairs| repair]] section of switch assembly [[260-1971-00]].
Refer to the the pictures and the [[260-1971-00/Repairs| repair]] section of switch assembly [[260-1971-00]].
==R10 Repair==
Fault pattern: U AC/DC and resistance (exept the lowest range) are 2% to low.
Fault: R10. Some resistors are to low (900 kΩ and 90 kΩ). The "stuff" inside has become liquid. The liquid has a low resistance.
Solution: Open R10. Clean R10 inside with isopropanol and in a ultrasonic bath.


==See Also==
==See Also==

Revision as of 10:59, 2 September 2023

Switch Repair

The contact bar in switches can break. Replacing the contact bar and contacts can be done without desoldering the switches. Refer to the the pictures and the repair section of switch assembly 260-1971-00.

R10 Repair

Fault pattern: U AC/DC and resistance (exept the lowest range) are 2% to low.

Fault: R10. Some resistors are to low (900 kΩ and 90 kΩ). The "stuff" inside has become liquid. The liquid has a low resistance.

Solution: Open R10. Clean R10 inside with isopropanol and in a ultrasonic bath.

See Also

Pictures