36,375
edits
m (Text replacement - "Category:1 MOhm Oscilloscope Probes" to "Category:1 MΩ Oscilloscope probes") |
No edit summary |
||
(8 intermediate revisions by 3 users not shown) | |||
Line 1: | Line 1: | ||
{{Probe Sidebar | |||
[[File:P6055 2.jpg | thumb | | |manufacturer=Tektronix | ||
The '''Tektronix P6055''' is a 10× probe pair for differential amplifiers such as e.g. the [[7A13]] having 1 MΩ Inputs and 20 to 47 pF input capacity. | |model=P6055 | ||
|summary=10× probe for differential applications | |||
|image=P6055 1.jpg | |||
|caption=Tektronix P6055 | |||
|for=1 MΩ scope inputs | |||
|introduced=1971 | |||
|discontinued=(?) | |||
|manuals= | |||
* [[Media:070-1115-00.pdf|Tektronix P6055 Manual]] | |||
* [[Media:Tek_p6055_1971_cat.pdf|P6055 in 1971 Catalog]] | |||
}} | |||
[[File:P6055 2.jpg | thumb | 250px | right]] | |||
The '''Tektronix P6055''' is a 10× probe pair for differential amplifiers such as e.g. the [[7A13]] and [[5A22N]] having 1 MΩ Inputs and 20 to 47 pF input capacity. | |||
The probe's resistive attenuation ratio is adjustable to an exact ×10 for the given amplifier input. | The probe's resistive attenuation ratio is adjustable to an exact ×10 for the given amplifier input. | ||
This allows precise gain matching to improve CMRR. | This allows precise gain matching to improve CMRR. | ||
Line 9: | Line 22: | ||
The P6055 has a [[readout pin]]. | The P6055 has a [[readout pin]]. | ||
{{ | {{BeginSpecs}} | ||
{{Spec | Attenuation | Adjustable to ×10 }} | |||
{{Spec | Input impedance | 1 MΩ // 10-12.5 pF}} | |||
{{Spec | Step response | < 4 ns}} | |||
{{Spec | Compensation range | 1 MΩ//12-47 pF}} | |||
{{Spec | CMRR | with [[7A13]] or [[1A5]], 20,000:1 from DC – 1kHz, 100:1 at 20 MHz }} | |||
{{Spec | Max. input | ±500 V maximum common-mode }} | |||
{{EndSpecs}} | |||
==Links== | |||
{{Documents|Link=P6055}} | |||
{{MissingPics}} | |||
[[Category:1 MΩ Oscilloscope probes]] | [[Category:1 MΩ Oscilloscope probes]] | ||
[[Category:Passive Differential probes]] |