Line 53: Line 53:
be found with an access hole drilled for R235 (GAIN SWITCHING BALANCE) since this
be found with an access hole drilled for R235 (GAIN SWITCHING BALANCE) since this
is adjusted to remove the trace shift when the probe switches from 1× to 10× mode.
is adjusted to remove the trace shift when the probe switches from 1× to 10× mode.
To calibrate the amplifier, first check the power supply voltages.  They are floating.
Do not check with respect to ground!  When adjusting, watch for exposed line voltage
in the power supply, especially the fuse holder.  The power supply must be adjusted
with everything connected together.  Note that you can check the voltages from the
back of the amplifier output board.


==Manuals==
==Manuals==
There are three versions of the manual
* 070-0756-00 - (C) 1968
* 070-7129-00 - (C) 1988
* 070-7129-01 - (C) 1992
The 1968 version assumes you have a [[1A5]] for absolute gain calibration of the probe
without the amplifier.


* [http://w140.com/Tektronix_P6046_Instruction_Manual.pdf P6046 Manual (no schematics, PDF)]
* [http://w140.com/Tektronix_P6046_Instruction_Manual.pdf P6046 Manual (no schematics, PDF)]