067-0587-02: Difference between revisions

Jump to navigation Jump to search
internals
No edit summary
(internals)
Line 11: Line 11:
}}
}}
The '''Tektronix 067-0587-02''' is a signal standardizer (calibration) plug-in for [[7000-series scopes]].
The '''Tektronix 067-0587-02''' is a signal standardizer (calibration) plug-in for [[7000-series scopes]].
It is a complete redesign of the [[067-0587-01]] standardizer.  It provides a calibrated staircase at 1 line/Div in Gain mode, a fast-rising pulse in Step Resp mode, an input for a CW signal to test bandwidth, and an input for a common mode test signal.
It is a complete redesign of the [[067-0587-01]] standardizer.  It provides a calibrated staircase at 1 line/Div in Gain mode, a fast-rising pulse in Step Resp mode, an input for a CW signal to test bandwidth, and an input for a common mode test signal. The 067-0587-02 was designed by [[Art Metz]].


{{BeginSpecs}}
{{BeginSpecs}}
Line 29: Line 29:


==Internals==
==Internals==
Internal signals are derived from a 1 MHz oscillator with selectable decade divisions down to 10 Hz repetition.
The clock is delayed, amplified and limited, then fed to a Schittky-diode based pulse shaper with L-C compensating networks. The output pulse has a rise time of max. 150 picoseconds with less than 2% aberrations.
The staircase generator for Gain mode is implemented using a 4 bit count and multiplexer.  The 0-volt step (No. 6) is held for five counts, producing a brighter trace to allow easy alignment wit the center of the screen.
For bandwidth measurements, there is a leveler circuit with a differential diode detector controlling an amplifier's gain in an AGC loop.


The 067-0587-02 is based on a number of Tektronix custom ICs from the 1 GHz [[7104]] scope generation, such as the [[155-0160-00|155-0160-00 trigger amplifier + source select chip]] and the [[155-0175-00|155-0175-00 trigger amplifier chip]].
The 067-0587-02 is based on a number of Tektronix custom ICs from the 1 GHz [[7104]] scope generation, such as the [[155-0160-00|155-0160-00 trigger amplifier + source select chip]] and the [[155-0175-00|155-0175-00 trigger amplifier chip]].
Line 35: Line 42:


==Links==
==Links==
* [http://amplifier.cd/Test_Equipment/Tektronix/Tektronix_7000_series_special/calibration-fixture.htm 067-0587-02 page @ amplifier.cd] (comparing -01 and -02, lots of pictures, text in German)
* [http://amplifier.cd/Test_Equipment/Tektronix/Tektronix_7000_series_special/calibration-fixture.htm 067-0587-02 @ amplifier.cd] (comparing -01 and -02, lots of pictures, text in German)
* TekScope Vol 11 No. 1, 1979, p.14.


==Pictures==
==Pictures==

Navigation menu