AM503: Difference between revisions

Jump to navigation Jump to search
386 bytes added ,  11 September 2015
no edit summary
No edit summary
No edit summary
Line 3: Line 3:
* [http://w140.com/am503a_service.pdf Tektronix AM503A Manual (PDF)]
* [http://w140.com/am503a_service.pdf Tektronix AM503A Manual (PDF)]
}}
}}
It works with the [[A6302]] or [[A6303]] current probes.
Versions AM503A and AM503B have digital readout of deflection factor.


{{BeginSpecs}}
{{BeginSpecs}}
{{Spec | Bandwidth |
{{Spec | Bandwidth |
* A6302: DC to 50 MHz
* A6302: DC to 50 MHz
* A6303: DC to 15 MHz }}
* A6312: DC to 100 MHz
* A6303: DC to 15 MHz
* A6304XL: DC to 2 MHz
}}
{{Spec | Sensitivity |  
{{Spec | Sensitivity |  
* A6302: 1 mA/Div to 5 A/Div
* A6302, A6302XL, A6312: 1 mA/Div to 5 A/Div
* A6303: 10 mA/Div to 50 A/Div }}
* A6303, A6303XL: 10 mA/Div to 50 A/Div  
}}
{{Spec | Range |
{{Spec | Range |
* A6302: max. 20 A continuous, 50 A peak
* A6302: max. 20 A continuous, 50 A peak
* A6303: max. 100 A continuous, 500 A peak
* A6303: max. 100 A continuous, 500 A peak
* A6304XL: max. 500 A continuous, 700 A peak
}}
}}
{{Spec | Scope output | 10 mV/Div}}
{{Spec | Scope output | 10 mV/Div}}
{{Spec | Power | 17 W}}
{{Spec | Power | 17 W}}
{{Spec | Compatible probes |
{{Spec | Compatible probes |
* [[A6302]] (20 A, DC-50 MHz)
* [[A6302]] (20/100 A, DC-50 MHz), [[A6302XL]] (20/100 A, DC-17 MHz)
* [[A6312]] (20 A, DC-100 MHz)
* [[A6312]] (20/100 A, DC-100 MHz)
* [[A6303]] (100 A, DC-15 MHz)
* [[A6303]] (100/500 A, DC-15 MHz), [[A6303XL]] (100/500 A, DC-10 MHz)
* [[A6304XL]] (500/700 A, DC-2 MHz)
* [[CT4]] with A6302, A6302XL or A6312 (2000/20000 A)
}}
}}
{{EndSpecs}}
{{EndSpecs}}


==Links==
==See also==
* [[AM5030]]
* http://www.tek.com/datasheet/current-measurement-system-probes


==Pictures==
==Pictures==

Navigation menu