A6901: Difference between revisions

Jump to navigation Jump to search
545 bytes added ,  30 January 2016
no edit summary
No edit summary
No edit summary
Line 1: Line 1:
The Tektronix A6901 Ground Isolation Monitor.
[[File:A6901.jpg | thumb | 350px | right]]
The '''Tektronix A6901 Ground Isolation Monitor''' connects between mains power and test equipment
without connecting protective ground, in order to safely allow referencing measurements to a low voltage
instead of ground, or break ground loops.
 
It monitors the floating equipment ground voltage. If the voltage exceeds 40 V peak,
or ground current exceeds a selectable limit of 0.5, 3.5 or 5 mA, the A6901 disconnects
mains power and grounds the test equipment.
 
{{MissingSpecs}}
 
==Manuals==


* [http://w140.com/tek_a6901.pdf Tektronix A6901 Manual (PDF)]
* [http://w140.com/tek_a6901.pdf Tektronix A6901 Manual (PDF)]


[[File:A6901 principle.jpg | thumb | 350px]]


[[Category:Amplifiers]]
[[Category:Power supplies]]