==Variable control==
Through an internal jumper, the front-panel variable control (red knob) can be selected to vary either the main sweep, the delayed sweep or the main sweep hold-off.  See i.a. [[Media:Tekscope 1975 V7 N5.pdf | TekScope Vol. 7 No. 5, 1975]] p.15+: Dave McCullough, ''Delayed gate aids oscilloscope digital measurements''.


==X-Y operation==
==X-Y operation==
The 7B53 supports X-Y mode by selecting the "Ampl" position on the Time/Div switch.  The X signal can be picked up from a vertical plug-in via the internal trigger path or applied via the "Main Trig In or Amp In" input, as determined by the Trigger Source switch setting. Note that the x10 magnifier affects horizontal deflection in X-Y mode.
The 7B53 supports X-Y mode by selecting the "Ampl" position on the Time/Div switch.  The X signal can be picked up from a vertical plug-in via the internal trigger path or applied via the "Main Trig In or Amp In" input, as determined by the Trigger Source switch setting. Sensitivity on the external input is 100 mV/Div,  which can be increased to 10 mV/DIv by selecting the x10 magnifier, or decreased to 1 V/Div by pushing the EXT÷10 button.