475: Difference between revisions

Jump to navigation Jump to search
467 bytes added ,  19 July 2017
no edit summary
No edit summary
No edit summary
Line 20: Line 20:
The optional [[DM44]] multimeter attaches to the top of the instrument.
The optional [[DM44]] multimeter attaches to the top of the instrument.


==Specifications==
{{BeginSpecs}}
[[Category:Specifications needed]] ''please add''
{{Spec | Bandwidth | 200 MHz (475), AC cutoff 10 Hz, switchable BW limit 20 MHz }}
{{Spec | Rise time | 1.75  ns (475) }}
{{Spec | Deflection | 2 mV/Div to 5 V/Div, 1-2-5 }}
{{Spec | Cascaded mode | 400 μV/Div, 50 MHz with CH1 input connected to CH2 VERT SIG OUT }}
{{Spec | Time base |
{{Spec | Input impedance | 1 MΩ // 20 pF }}
{{Spec | Triggering | 0.3 Div (int) or 50 mV (ext) to 40 MHz, increasing to 1.5 Div/250 mV at 200 MHz; AC coupling >60 Hz; LF REJ >50 kHz, HF REJ <50 kHz }}
{{EndSpecs}}


==Internals==
==Internals==

Navigation menu