The '''Tektronix TDS5034''' is a 350 MHz, 5 GS/s "digital phosphor" DSO.
{{Oscilloscope Sidebar |
title=Tektronix TDS5034 |
summary=Digital Storage 350 MHz 4-channel scope |
image= Tek tds5034b 1.jpg |
caption=Tektronix TDS5034 |
introduced=? |
discontinued=? |
manuals=
* [[Media: 55W_14869_15_TDS5000_Series_DPO_Datasheet.pdf| TDS 5000 Series DPO Datasheet]] (743 KB, PDF)
* [[Media: 071-0876-02_TDS5000_Series_DPO_User_Manual.pdf| TDS5000 Series DPO User Manual]] (5.9 MB, PDF)
* [[Media: 071-1003-00_TDS5104,_TDS5054,_TDS5052_Read_This_First_Manual.pdf| TDS5000 Read This First Manual]] (1.1 MB, PDF)
* [[Media: 071-1004-02_TDS5000_Series_DPO_Service_Manual.pdf| TDS5000 Series DPO Service Manual]] (3.6 MB, PDF)
* [[Media: 071-1316-00_TDS5032_&_TDS5034_Supplement_Manual.pdf| TDS5032 & TDS5034 Supplement Manual]] (144 KB, PDF)
}}


==Manuals==
The '''Tektronix TDS5034''' is a four-channel, 350 MHz, 5 GS/s digital phosphor oscilloscope with a 10.4 in. LCD active-matrix color display.  This series of scopes combines 100,000 wfms/s continuous waveform capture rate and real-time intensity grading so you can see the problem and solve it. An optional touch screen was available for all models but was standard on the TDS5054B and TDS5104B.  The operating system is Windows 2000.
* [http://w140.com/service-manual-tds5034b-tds5054b-tds5104b.pdf Tektronix TDS5000B Series Service Manual (PDF)]
 
Besides the channel inputs, the AUX IN (auxiliary trigger), and Floppy drive, the Left Side panel has two USB ports, two PS-2 ports (Keyboard, Mouse), two VGA connectors, one Parallel port, GPIB, RJ-45, Serial port, line out, line in, and microphone in, and a CD-ROM drive (Opt 16 replaces it with a CD-RW drive), and the Rear panel has a EXT REF (time base reference input), AUX OUT (auxiliary trigger pulse output), and SIGNAL OUT (buffered CH 3 signal on the TDS5054 and TDS5104 only).
 
Applications Modes available are FastAcq Acquisition (optimizes the instrument for analysis of dynamic signals and capture of infrequent events), Sample, Peak Detect (captures narrow glitches <1 ns at all real-time sample rates), Averaging (from 2 to 10,000 waveforms included in average), Envelope (from 2 to 10,000 waveforms included in min-max envelope), Hi-Res (real-time boxcar averaging reduces random noise and increases resolution), Waveform Database (accumulates a waveform database that provides a three-dimensional array of amplitude, time, and counts), and FastFrame Acquisition (acquisition memory divided into segments, maximum trigger rate >100,000 waveforms per second).
 
There are many options for signal acquisition, measurement, and analysis.  The Overview section of the User Manual is a great place to become familiar with the many functions available in the use of this sophisticated instrument.
 
{{BeginSpecs}}
{{Spec | Bandwidth and Channels| 350 MHz; Bandwidth Limit switchable between 20 MHz and 150 MHz; 4 Channels }}
{{Spec | Vertical Sensitivity| 1 mV to 10 V/div (1 MΩ), 1 mV to 1 V/div (50 Ω) }}
{{Spec | Time Base Range| 200 ps to 1000 s/div }}
{{Spec | Sample Rate (Max)| 5 GS/s on 1 channel, 2.5 GS/s on 2 channels, 1.25 GS/s on 3-4 channels }}
{{Spec | Max Record Length per channel|
Std mem: 400 k (1 ch), 200 k (2 ch) , 100 k (4 ch); Opt 1M mem: 2 M (1 ch), 1 M (2 ch), 500 k (4 ch); Opt 2M mem: 8 M (1 ch), 4 M (2 ch), 2 M (4 ch) }}
{{Spec | Max Record Length per channel (B series)|
Std mem (TDS5034B): 8 M (1 ch), 4 M (2 ch), 2M (4 ch); Opt 3M mem (TDS5034B): 16 M (1 ch), 8 M (2 ch) , 4 M (4 ch) }}
{{Spec | Waveform Measurements| Arithmetic: add, subtract, multiply, and divide waveforms;, Fast Fourier Transform (FFT): Magnitude; Vertical Units: Magnitude, Linear, dB, dBm; Window Functions: Rectangular, Hamming, Hanning, Kaiser-Bessel, Blackman-Harris, Gaussian, Flattop2, Tek Exponential; OPTIONAL Math: Algebraic Expressions: Extensive algebraic expressions including waveforms, scalars, and results of parametric measurements; Calculus: Integrate, differentiate; Frequency Domain Functions: Spectral magnitude and phase, real and imaginary spectra; Additional Vertical Units: Phase, degrees, radians }}
{{Spec | Probes| Four [[P5050]] 500 MHz ×10 probes, [[BNC connector]] with a [[readout pin]] }}
{{Spec | Power| 100 − 240 V<sub>AC</sub>, 47 − 63 Hz, CAT II, <220 W }}
 