The '''Sony/Tektronix 3086''' is a 3 GHz Real Time Spectrum Analyzer.  
{{Oscilloscope Sidebar |
title=Sony/Tektronix 3086 |
summary=Spectrum Analyzer 10 MHz to 3 GHz |
image= Tektronix_3086_Waveform_image.jpg |
caption=Sony/Tektronix 3086 |
introduced=1996*|
discontinued=(?) |
manuals=
Need User Manual
* [[Media: Tek 3086 3066 071050101.pdf| Sony-Tek 3066_3086 User Manual (PDF)]]
* [[Media: 071-0502-01_Sony-Tek_3066_3086_Programmer Manual.pdf| Sony-Tek 3066_3086 Programmer Manual (PDF)]]
* [[Media: Tektronix_3066_3086_Catalog_Datasheet.pdf| Tektronix 3066_3086 Catalog Datasheet (PDF)]]
}}


The [[3066]] is the same without the wideband IF module and I/Q inputs.
The '''Sony/Tektronix 3086''' is a portable real time spectrum analyzer with a frequency range of 10 MHz to 3 GHz in coax.  This spectrum analyzer is equipped with a 3 GHz down converter for analyzing radio frequency (RF) signals and allows seamless capture of frequency spectra with spans of up to 30 MHz.  This allows the spectrum of burst signals, such as those from GSM, IS-136, PDC or partial rate CDMA mobile stations that are captured and examined.


The instruments are based on a Windows 95 computer with a 2 GB disk.
The 3026 can obtain time domain and frequency domain data simultaneously through a new architecture which allows capturing a block of frequencies all at once.  Because this instrument samples these full frames constantly (rather than waiting for each discrete frequency step to be measured), the signal can come or go as it pleases and the real-time spectrum analyzer will detect the change instantly.


{{MissingSpecs}}
A frequency mask trigger allows capture of randomly occurring or infrequent signals such as intermittent spurious emissions, as well as, the capture of “burst-on” periods, thus maximizing the number of bursts that can be captured in memory for subsequent spectral, time, or modulation analysis.  The user can graphically define the frequency trigger mask; this allows the trigger condition to be generated by a particular event in the frequency domain.


==Manuals==
Waveforms and measurements are displayed on the 12.1 inch full (256 max) colors (1024 x 768) TFT display which allows 1, 2, or 4 views of spectrum, waterfall displays, spectrograms, and digital modulation analysis screens. Constellation and vector diagrams can be displayed as well as frequency, phase, magnitude, I and Q versus time for in-depth analysis of digitally modulated signals.
* [[Media:Tek 3086 3066 071050101.pdf|Tektronix 3066 and 3086 User Manual (no schematics, PDF)]]


{{MissingPics}}
Users can move a marker through the time record and analyze the modulation of any point.  In particular, signals with symbol rates up to 20.48 Msymbols/sec may be analyzed during and after capture.  This facilitates analysis of W-CDMA signals with chip rates up to 16.384 Mcps.  Data is stored and retrieved from the built in hard drive (2.1 GB) and 3.5 inch floppy disk drive.


Front panel inputs/outputs are: RF INPUT 50Ω [[N connector]]; EXT TRIG [[BNC connector]]; and [[Floppy]] disk drive.


[[Category:Spectrum Analyzers]]
Rear panel inputs/outputs are: PRINTER female [[DB25 connector]]; SCSI female [[SCSI connector]]; 10 MHz REF OUTPUT [[BNC connector]]; 10 MHz REF INPUT [[BNC connector]]; I INPUT [[BNC connector]]; and Q INPUT [[BNC connector]].
 
{{BeginSpecs}}
{{Spec | Frequency Range | RF Mode: 10 MHz to 3 GHz; Baseband Mode: DC to 10 MHz }}
{{Spec | Frequency Span Range | RF Mode: 100 Hz to 3 GHz; Baseband Mode: 100 Hz to 6 MHz }}
{{Spec | RF Input impedance | 50 Ω }}
{{Spec | Maximum Safe Input Power | +30 dBm }}
{{Spec | Input Modes | RF, Baseband }}
{{Spec | Attenuator Range (used in Baseband Mode only) | (Spec missing from datasheet) }}
{{Spec | Reference Level Range | RF Mode: -50 dBm to +30 dBm; Baseband Mode: -30 dBm to +30 dBm }}
{{Spec | Spectral Frame Update Interval | 1024-point: 6400 ms to 80 µs; 256-point: 1600 ms to 20 µs }}
{{Spec | Triggering Modes | Auto, Normal [RF (span ≤ 6MHz) and Baseband Input, Wideband and IQ Input], Quick, Delayed, Interval, Quick-interval, and Timeout }}
{{Spec | Trigger Source | Internals, External }}
{{Spec | Frequency Trigger Mask | Resolution: 1 bin; Level Range: 0 dBfs to -79 dBfs }}
{{Spec | Time Trigger Mask | Resolution: 1 data point; Level Range: 0 dBfs to -40 dBfs }}
{{Spec | Marker | Types: Normal, Delta, Band-Power; Search Function: Peak Right, Peak Left, Maximum }}
{{Spec | Measurement Function | Noise Power, Power with Band, C/N, Adjacent Channel Power, Occupied Bandwidth }}
{{Spec | Digital Zoom | Zoom Ratio: 2 to 1000; Max Span in Zoom Mode: 30 MHz (Wideband, IQ) }}
{{Spec | Input Equivalent Noise | RF: −140 dBm/Hz; Baseband: -150 dBm/Hz; Wideband: -140 dBm/Hz }}
{{Spec | Spurious Free Dynamic Range | RF: 80 dB at 50 kHz span; 65 db at 1/2/5 MHz span; Baseband: 85 db at 50 kHz; 70 db at 1/2/5 MHz span; Wideband/IQ: 65 dB at 10 MHz span; 60 db at 20 MHz and 30 MHz span }}
{{Spec | Included Accessories | User Manual, U.S. Power Cord, Mouse }}
{{Spec | Weight | 28 kg (62 lbs) }}
{{Spec | Power | 100 − 240 V<sub>AC</sub>, 47 − 66 Hz; 250 W max }}
{{Spec | Operating Atmospherics | Temperatures: +5 °C to +40 °C;  Humidity: 80 %; Altitude: Up to 3 km (10,000 ft) }}
{{EndSpecs}}
 
==Options==
Opt. 1S: Data capture software.
Opt. C3: Three years of calibration services.
Opt. C5: Five years of calibration services.
Opt. D1: Test Data.
Opt. D3: Test Data (requires Opt. C3)
Opt. R3: Three years of repair protection.
 
==Recommended Accessories==
PC Display Analysis Software - 3086: Order SL7PC86
 
==Pictures==
<gallery>
Sony-Tek 3086_Waveform_image.jpeg
Sony-Tek 3086_1.jpg
Sony-Tek 3086_2.jpg| Same as 3066 but with I/Q Input connectors instead of Sync In/Out connectors
Sony-Tek 3086_3.jpg
Sony-Tek 3086_4.jpg
Sony-Tek 3086_5.jpg
Sony-Tek 3086_6.jpg
Sony-Tek 3086_7.jpg
</gallery>
 
[[Category: Spectrum Analyzers]]
* Year of introduction confirmed by Dave Brown (VintageTek Museum) with Sony-Tek Product Planner, Koichi Sega.  First year shown in Tek Catalog year 2000 when relationship with Advantest Corporation ended in 1999.
1,192

edits

Navigation menu