{{TM500 | mfg=Tektronix | type=DP501 | function=1.3 GHz ÷ 16 prescaler | class=digital function | image=Tek dp501.jpg | introduced=1984 | discontinued= | {{TM500 | mfg=Tektronix | type=DP501 | function=1.3 GHz ÷ 16 prescaler | class=digital function | image=Tek dp501.jpg | introduced=1984 | discontinued=1993 | | ||
manuals= | manuals= | ||
* [[Category:Manual needed]] ''Manual needed'': [[Media:070-4332-00.pdf|DP501 manual (PDF)]] | |||
* [[Media:DP501-specs.pdf|DP501 specifications (PDF)]] | |||
* [[Media:TM-9-6695-275-34P.pdf|Repair Parts and Special Tools List (TM 9-6695-275-34P)]] | * [[Media:TM-9-6695-275-34P.pdf|Repair Parts and Special Tools List (TM 9-6695-275-34P)]] | ||
* [[Media: | * [[Media:ServiceTekNotes 40 Aug 1985.pdf|ServiceTekNotes 40, Aug 1985, p.2]]: DC503A/DP501 interfacing instructions | ||
}} | }} | ||
It is specifically designed to interface with the [[DC503A]], [[DC509]], [[DC5009]], [[DC510]] and [[DC5010]] Universal Counter Timers through the TM500 rear interface to make the counter display the correct frequency directly. This is done by routing the counter's reference clock to the DP501, where it is also divided by 16, and fed back to the counter as the new reference, or through a dedicated prescaler signal for microprocessor based counters like the DC50(0)9. | |||
{{ | {{BeginSpecs}} | ||
{{Spec | Input Frequency | 100 MHz to ≥1.3 GHz }} | |||
{{Spec | Sensitivity | ≤20 mV<sub>RMS</sub> (−21 dBm) from 100 MHz to 1 GHz; ≤30 mV<sub>RMS</sub> (−17 dBm) from 1 GHz to 1.3 GHz }} | |||
{{Spec | Input Impedance | 50 Ω, AC coupled; VSWR ≤2.2 : 1 }} | |||
{{Spec | Output | ≥200 mV<sub>p-p</sub> into 50 Ω, from a 50 Ω source }} | |||
{{Spec | Max. Input | +20 dBm (damage level ≥ +25 dBm) }} | |||
{{Spec | Power | 19 VA }} | |||