The '''Tektronix 067-0500-00''' (also known as [[3M1]]) is a calibration fixture for adjusting
{{Plugin Sidebar
the normalizing capacitor behind the plug-in connector in [[560-series scopes]].
|manufacturer=Tektronix
|series=560-series scopes
|type=067-0500-00
|summary=Calibration fixture
|image=067-0500-00_front.jpg
|caption=067-0500-00 Front
|introduced=(?)
|discontinued=(?)
|designers=
|manuals=
* [[Media:062-0806-01.pdf|Tektronix 067-0500-00 Manual (-01) (PDF)]]
* [[Media:062-0806-00.pdf|Tektronix 067-0500-00 Manual (-00) (PDF, OCR)]]
}}
The '''Tektronix 067-0500-00''' (also known as [[3M1]]) is a calibration fixture for adjusting the normalizing capacitor behind the plug-in connector in [[560-series scopes]].


It consists of a Schmitt trigger that is driven from  
It consists of a Schmitt trigger that is driven from a single-ended input coming from a BNC connector on the front panel.   
a single-ended input coming from a BNC connector on the front panel.   
The Schmitt trigger produces balanced output signals with controlled source impedance.   
The Schmitt trigger produces balanced output signals with controlled source impedance.   
The normalizing capacitor is adjusted using a procedure similar to adjusting
The normalizing capacitor is adjusted using a procedure similar to adjusting probe compensation.   
probe compensation.  When adjusted correctly, the voltage division ratio formed
When adjusted correctly, the voltage division ratio formed by the plug-in's output impedance and the mainframe's input impedance should be a real value which does not change across the specified frequency range of the scope.
by the plug-in's output impedance and the mainframe's input impedance should be
After the scope is normalized using the 067-0500-00, other plug-ins can be adjusted to produce clean transient response in the normalized scope.
a real value which does not change across the specified frequency range of the scope.
After the scope is normalized using the 067-0500-00, other plug-ins can be adjusted
to produce clean transient response in the normalized scope.
 