36,375
edits
No edit summary |
No edit summary |
||
Line 17: | Line 17: | ||
Optional 10× or 100× attenuator tips (010-0364-00, 010-0365-00) provide extended input voltage range; the optional [[VP-2]] voltage pickoff converts it to a 50 Ω feed-through sampling head. | Optional 10× or 100× attenuator tips (010-0364-00, 010-0365-00) provide extended input voltage range; the optional [[VP-2]] voltage pickoff converts it to a 50 Ω feed-through sampling head. | ||
{{BeginSpecs}} | |||
{{Spec | Rise time | ≤350 ps (×10, ≤400 ps; ×100, ≤500 ps) }} | |||
{{Spec | Input impedance | 100 kΩ // 2.3 pF (×10, 1 MΩ // 2 pF; ×100, 1 MΩ // 1.7 pF) }} | |||
{{Spec | Maximum signal | 2 V<sub>p-p</sub> (×10, 20 V<sub>p-p</sub>; ×100, 200 V<sub>p-p</sub>) }} | |||
{{Spec | Maximum safe input | max. ±100 V, max. sine input 20 V<sub>p-p</sub> (×10 or ×100, max. ±350 V, max. sine input 200 V<sub>p-p</sub>) }} | |||
{{Spec | Features | | |||
* no trigger pickoff for internal triggering | |||
* offset voltage of either ±1 V or ±2 V (at ×1) can be selected via a toggle switch on the front panel | |||
}} | |||
{{EndSpecs}} | |||
==Differences between S-3 and S-3A == | |||
The S-3A sampling head has a 10× improvement in baseline shift due to cable flexing. | |||
== Differences between S-3 and S-3A == | |||
The S-3A sampling head has | |||
All other specifications are identical to the ordinary S-3. | All other specifications are identical to the ordinary S-3. | ||
Line 38: | Line 35: | ||
but some changes are also made to the circuits in the body of the probe. | but some changes are also made to the circuits in the body of the probe. | ||
== Difference between the S-3 and the P6038 == | ==Difference between the S-3 and the P6038 == | ||
The S-3 probe contains a compensated attenuator before the sampling bridge. | The S-3 probe contains a compensated attenuator before the sampling bridge. | ||
The [[P6038]] does not. | The [[P6038]] does not. | ||
The S-3 probe has a JFET buffer between the holding cap and the probe cable. | The S-3 probe has a JFET buffer between the holding cap and the probe cable. | ||
The P6038 does not. | The P6038 does not. | ||
Both use the [[152-0368-00]] sampling bridge. | Both use the [[152-0368-00]] sampling bridge. The sampling pulse is generated by a [[snap-off diode]] ([[152-0368-00]]). | ||
== Repair of Sampling Diodes == | == Repair of Sampling Diodes == | ||
Line 49: | Line 46: | ||
==Pictures== | ==Pictures== | ||
'''S-3''' | |||
<gallery> | |||
Tek s3 probe sample bridge.jpg|S-3 probe sampling bridge | |||
Tek s3 sample bridge2.jpg|S-3 probe sampling bridge | |||
S3_frontend.png | S-3 frontend circuit | |||
</gallery> | |||
'''S-3A''' | |||
<gallery> | <gallery> | ||
S3a attenuator.jpg|010-0364-00 ×10 attenuator for S-3A | S3a attenuator.jpg|010-0364-00 ×10 attenuator for S-3A | ||
S3a box front.jpg|Front | S3a box front.jpg|Front | ||
Line 58: | Line 62: | ||
S3a strobe board.jpg|S-3A strobe board | S3a strobe board.jpg|S-3A strobe board | ||
S3a probe_internal_bothsides.jpg|S-3A probe both sides | S3a probe_internal_bothsides.jpg|S-3A probe both sides | ||
</gallery> | </gallery> | ||
[[Category:7000 and 3S series sampling heads]] | [[Category:7000 and 3S series sampling heads]] |