* If no LED or display segment is lit after trun on, check for +15 V from U500 Voltage regulator IC type 723.
* If no LED or display segment is lit after trun on, check for +15 V from U500 Voltage regulator IC type 723.
* All ICs are placed in IC sockets. Contact problems can occur. Removing and inserting ICs helps.
* All ICs are placed in IC sockets. [[bad TI IC sockets|Contact problems can occur]]. Removing and inserting ICs helps.
* Check board is mounted to frame with plastic washers in between near pcb conductor tracks, otherwise a short to GND can occur.