{{Spec|Temperature|–200 °C to +630 °C using a PT-100 (DIN 43760) RTD sensor, 4-wire or 3-wire connection }}
{{Spec|Temperature|–200 °C to +630 °C using a PT-100 (DIN 43760) RTD sensor, 4-wire or 3-wire connection }}
{{Spec|Reading rate|3½ digits: 3.3 ms integration, 76 rdgs/s; 4½ digits: 16.66 ms integration, 36 rdgs/s; 5½ digits: 100 ms integration, 9 rdgs/s }}
{{Spec|Reading rate|3½ digits: 3.3 ms integration, 76 rdgs/s; 4½ digits: 16.66 ms integration, 36 rdgs/s; 5½ digits: 100 ms integration, 9 rdgs/s }}
{{Spec|Power|105–125 V or 210–250 V, 50–400 Hz, 24 VA  (with low-voltage transformer option, 90–110 V or 180-220 V) }
{{Spec|Power|105–125 V or 210–250 V, 50–400 Hz, 24 VA  (with low-voltage transformer option, 90–110 V or 180-220 V) }}
{{Spec|Connections|Analog: 5-way binding posts on front and back, switch-selectable; Digital: Trigger Input and Voltmeter Complete output, rear panel, [[BNC connector]]s, TTL level; GPIB }}
{{Spec|Connections|Analog: 5-way binding posts on front and back, switch-selectable; Digital: Trigger Input and Voltmeter Complete output, rear panel, [[BNC connector]]s, TTL level; GPIB }}
{{EndSpecs}}