PG505: Difference between revisions

Jump to navigation Jump to search
no edit summary
No edit summary
No edit summary
Line 8: Line 8:
|discontinued=1983
|discontinued=1983
|manuals=
|manuals=
* [[Media:070-1583-01.pdf|Tektronix PG505 Manual 070-1583-01]]
* [[Media:070-1583-01.pdf|Tektronix PG505 Manual 070-1583-01]] (OCR)
: '''''See erratum'''''
: '''''See errata'''''
}}
}}
The output section is isolated from chassis ground and can float up to ±200 V.
The output section is isolated from chassis ground and can float up to ±200 V.
Line 28: Line 28:
{{PatentLinks|PG505}}
{{PatentLinks|PG505}}
===Notes===
===Notes===
* '''''Manual erratum:''' In manual 070-1583-01, Fig. 4.3 Output Delay Range Check (p. 4-5), the PG 505 connections are drawn correctly but the connector labels are swapped – the left connector fed from the FG501A via the BNC "T" is the Delay/Duration input, not the output.''
* '''''Manual errata:''' In manual 070-1583-01, Fig. 4.3 Output Delay Range Check (p. 4-5), the PG 505 connections are drawn correctly but the connector labels are swapped – the left connector fed from the FG501A via the BNC "T" is the Delay/Duration input, not the output.  In Table 5-2, Rear Connector Pin Assignments, B7 is incorrectly labeled –33.5 V, it is the NPN pass transistor's collector.''
* The PG505 is shipped with the output connector ground connected to chassis ground through a removable ground strap.  This must be cut/removed to float the output, in which case output common remains grounded through a 100 kΩ resistor (R545).  The manual says this resistor may be removed but doing so causes unspecified "degradation of performance".
* The PG505 is shipped with the output connector ground connected to chassis ground through a removable ground strap.  This must be cut/removed to float the output, in which case output common remains grounded through a 100 kΩ resistor (R545).  The manual says this resistor may be removed but doing so causes unspecified "degradation of performance".


Navigation menu