130: Difference between revisions
No edit summary |
No edit summary |
||
Line 13: | Line 13: | ||
{{BeginSpecs}} | {{BeginSpecs}} | ||
{{Spec | Ranges | 3, 10, 30, 100 or 300 pF / 3, 10, 30, 100 or 300 μH full scale}} | {{Spec | Ranges | 3, 10, 30, 100 or 300 pF / 3, 10, 30, 100 or 300 μH full scale}} | ||
{{Spec | Accuracy | 3% FS, | {{Spec | Accuracy | 3% FS, 1% FS with "careful" calibration using [[S-30|S-30 Delta Standard]]}} | ||
{{Spec | Measurement voltage | <1 V for capacitors, <0.25 V for inductors, 120-140 kHz }} | {{Spec | Measurement voltage | <1 V for capacitors, <0.25 V for inductors, 120-140 kHz }} | ||
{{Spec | Guard output | 250 Ω source impedance, can drive 200 pF }} | {{Spec | Guard output | 250 Ω source impedance, can drive 200 pF }} | ||
Line 22: | Line 22: | ||
* [http://bama.edebris.com/download/tek/130/Tek-130_LC_Meter.pdf Tektronix 130 Manual (PDF)] | * [http://bama.edebris.com/download/tek/130/Tek-130_LC_Meter.pdf Tektronix 130 Manual (PDF)] | ||
* [http://w140.com/tek_fcp/tek_type_130_factory_cal_proc.pdf Tektronix 130 Factory Calibration Procedure (PDF)] | * [http://w140.com/tek_fcp/tek_type_130_factory_cal_proc.pdf Tektronix 130 Factory Calibration Procedure (PDF)] | ||
==Links== | |||
* Q+A: Type 130 L-C Meter and S-30 Delta Standards. Service Scope No. 18, Feb 1963 | |||
<gallery> | <gallery> |
Revision as of 10:23, 11 September 2015
The Tektronix Type 130 is a self-contained instrument that measures inductance and capacitance. When measuring inductance, the 130 makes an LC oscillator using the device (inductance) under test and a capacitor internal to the 130. When measuring capacitance, the 130 makes an LC oscillator using the device (capacitance) under test and an inductor internal to the 130. The frequency of the resulting LC oscillator is measured by mixing it down using a fixed reference frequency, generating a pulse at each zero-crossing, and integrating (low-pass filtering) the pulse train. This produces a voltage that is proportional to the frequency difference between the LC oscillator and the reference oscillator. This voltage is displayed on a d'Arsonval (moving coil) meter on the front panel of the 130.
The Type 130 was designed by Cliff Moulton.
Key Specifications
Ranges | 3, 10, 30, 100 or 300 pF / 3, 10, 30, 100 or 300 μH full scale |
---|---|
Accuracy | 3% FS, 1% FS with "careful" calibration using S-30 Delta Standard |
Measurement voltage | <1 V for capacitors, <0.25 V for inductors, 120-140 kHz |
Guard output | 250 Ω source impedance, can drive 200 pF |
DUT connection | UHF connector (DUT) + 4 mm jack (guard) |
Manuals
Links
- Q+A: Type 130 L-C Meter and S-30 Delta Standards. Service Scope No. 18, Feb 1963
-
Early (Brown Era) Type 130
-
Early (Brown Era) Type 130
-
-
-
-
-
-
-
-
-
-