7000 Series plug-in interface: Difference between revisions

no edit summary
No edit summary
No edit summary
Line 125: Line 125:
* ''Sweep Inhibit'' −
* ''Sweep Inhibit'' −
* ''Delay Gate'' − +Gate Output on mainframe uses Delay Gate from H slot B when Delay Gate is selected
* ''Delay Gate'' − +Gate Output on mainframe uses Delay Gate from H slot B when Delay Gate is selected
* ''Single sweep ready'' −
* ''Single sweep ready'' − 0/+5 V, high when single sweep is ready
* ''X comp. inhibit'' − Connected to ground in amplifiers, switched to ground by timebase set to amplifier (X-Y) mode to enable X-Y delay compensation in scopes with that function ([[7904|7904 Opt.2]], [[7104|7104 Opt.2]])
* ''X comp. inhibit'' − Connected to ground in amplifiers, switched to ground by timebase set to amplifier (X-Y) mode to enable X-Y delay compensation in scopes with that function ([[7904|7904 Opt.2]], [[7104|7104 Opt.2]])
* ''Signal'' − Signal output from plugin to mainframe.  Differential, 50 Ω per side. 50 mV/Div (differential).