The '''Keithley 610C Solid State Electrometer''' can measure voltages (with a very high input resistance), very low currents, high resistances and charge.
The '''Keithley 610C Solid State Electrometer''' can measure voltages (with a very high input resistance), very low currents, high resistances and charge.


{{BeginSpecs}}
{{BeginSpecs}}
{{Spec|Voltage|11 ranges, 1 mV to 100 V full scale (× 1 / × 3 steps)}}
{{Spec|Voltage|1 mV to 100 V full scale in 11 ranges (×1 / ×3 steps) }}
{{Spec|Input resistance (volts mode)|>10<sup>14</sup> Ω or selectable 10 Ω to 10<sup>11</sup> Ω in decade steps}}
{{Spec|Input resistance (volts mode)|>10<sup>14</sup> Ω or selectable 10 Ω to 10<sup>11</sup> Ω in decade steps}}
{{Spec|Accuracy|1% of full scale, noise 25 μV}}
{{Spec|Accuracy|1% of full scale, noise 25 μV}}
{{Spec|Zero offset drift|<1 mV per day / 150 μV per °C}}
{{Spec|Zero offset drift|<1 mV per day / 150 μV per °C}}
{{Spec|Current|28  ranges, 10<sup>–14</sup> A to 0.3 A full scale}}
{{Spec|Current|10<sup>–14</sup> A to 0.3 A full scale in 28 ranges}}
{{Spec|Resistance|25 linear  ranges, 100 Ω to 10<sup>14</sup> Ω full scale}}
{{Spec|Resistance|100 Ω to 10<sup>14</sup> Ω full scale in 25 linear ranges}}
{{Spec|Charge|17 ranges, 10<sup>–13</sup> C to 10<sup>–3</sup> C (Amprere-seconds) full scale}}
{{Spec|Charge|10<sup>–13</sup> C to 10<sup>–3</sup> C (Amprere-seconds) full scale in 17 ranges}}
{{Spec|Display|Analog meter with mirror scale; positive, negative and center-zero display modes}}
{{Spec|Display|Analog meter with mirror scale; positive, negative and center-zero display modes}}
{{Spec|Power|105−125 V or 210−250 V, 50−60 Hz (some units labelled 50−1000 Hz), 10 W}}
{{Spec|Accessories|<small>
* Model 2503 Static Detector Probe (for measurements of charge on flat surfaces)
* Model 6101A Shielded Probe (shielded low-noise cable with a needle-point probe)
* Model 6103C Divider Probe (1000:1 voltage divider with a 3.8×10<sup>11</sup> input resistance)
* Model 6105 Resistivity Chamber (test fixture for measurement of surface and volume resistivity)
</small>}}
{{EndSpecs}}
{{EndSpecs}}


The unit has two sets of outputs on the back panel — one to attach a DVM or recorder, with ±3 V or ±1 mA full scale, using a special connector; the other output (marked "×1") tracks the input voltage. It is therefore possible to use the 610C as an amplifier, with a bandwidth of 40 kHz at ×1 down to 100 Hz at maximum amplification.
The unit has two sets of outputs on the back panel — one to attach a DVM or recorder, with ±3 V or ±1 mA full scale, using an [[Amphenol 80 series connectors|Amphenol 80-PC2F connector]], the other output (marked "×1") tracks the input voltage (in "Normal" mode only, s.b.). It is therefore possible to use the 610C as an amplifier, with a bandwidth of 40 kHz at ×1 down to 100 Hz at maximum amplification.


The input connector is a [[UHF connector|PL-259 ("UHF") socket]].  Newer electrometers use triaxial input connectors, with the inner shield (guard) driven by an amplifier to be at the same potential as the input, in order to minimize leakage currents. It is possible to use a guarded input connection on the 610C through an adapter, although the guard potential is only available on the back of the instrument, at the "× 1" output. (The terminal labelled "guard" is at input ground potential.)
The input connector is a [[UHF connector|PL-259 ("UHF") socket]].  Newer electrometers use triaxial input connectors, with the inner shield (guard) driven by an amplifier to be at the same potential as the input, in order to minimize leakage currents. It is possible to use a guarded input connection on the 610C through an adapter, although the guard potential is only available on the back of the instrument, at the "× 1" output. (The terminal labelled "guard" is at input ground potential.)


==Measurement Principle==
==Measurement Principle==
When set to the "Normal" feedback, the 610C measures currents through the voltage drop across a shunt resistor. The large range switch selects the shunt resistance in decade steps between 10 Ω and 10<sup>11</sup> Ω. Since only the voltage is measured, these shunts can also be used as defined input resistances in Volts mode (1 / current range), for example in the 1 nA (10<sup>-9</sup> A) range, a 1 GΩ resistance is connected across the input.
When set to the '''"Normal" feedback mode''', the 610C measures an input voltage, or an input current as the voltage drop across a shunt resistor.  
The large range switch selects the shunt resistance in decade steps between 10 Ω and 10<sup>11</sup> Ω, or leaves the input open.  


In "Fast" feedback mode, the current range resistor is connected between the amplifier output and the input, turning the instrument into a feedback amperemeter (transimpedance amplifier) and reducing the input voltage burden to less than 100 μV. (Modern pico/nanoamperemeters work this way.)
Because only the input voltage is measured, the shunts can also be used as defined input resistances in Volts mode (1 / current range).
For example, in the 1 nA (10<sup>-9</sup> A) range, a 1 GΩ resistance is connected across the input.
 
Note that in Normal mode, the "×1" output on the rear follows the input voltage within the full range, i.e. when the input is at +100 V, so is that output (for loads ≤100 μA).
[[File:Transimpedance amplifier.jpg|thumb|250px|right|Principle of transimpedance amplifier ("Fast" feedback mode in 610C)]]
 
In '''"Fast" feedback mode''', the current range resistor is connected between the amplifier output and the input, turning the instrument into a feedback amperemeter (transimpedance amplifier) and reducing the input voltage burden to less than 100 μV. (Modern pico/nanoamperemeters work this way.)


Because there is very little (change of) input voltage, the meter's input capacitance has no effect, and the meter responds faster, hence the name.
Because there is very little (change of) input voltage, the meter's input capacitance has no effect, and the meter responds faster, hence the name.
Line 40: Line 54:


==Internals==
==Internals==
At the core of the 610C is an operational amplifier with MOSFET inputs that is constructed from discrete transistors and works with +/- 120 V supplies, allowing it to measure voltages of up to 100 V in either polarity without needing the resistive input divider that is commonly found in electronic meters.
At the core of the 610C is an operational amplifier with MOSFET inputs that is constructed from discrete transistors and has a ±100 V range for input and output voltages.
This is done by bootstrapping – the output stage operates on ±120 V supplies, and the actual amplifier uses a floating ±9 V supply referred to that output.
As a result, the instrument can measure voltages of up to 100 V in either polarity without needing a resistive input divider, thereby preserving the MOSFET gate's high input resistance of >10<sup>14</sup> Ω over the entire ±100 V input range.


Note that the "× 1" output on the rear follows the input voltage within the full range, i.e. when the input is at +100 V, so is that output.
The opamp has a three-level zero adjustment, with switches for coarse (at the rear panel) and medium levels, and a 10-turn potentiometer for fine nulling.
Zero drift is fairly low once warmed up, especially compared to the predecessor units like the [[Keithley 610B]] which still relied on electrometer vacuum tubes in the input stage.


The opamp has a three-level zero adjustment, with switches for coarse and medium levels, and a 10-turn potentiometer for fine nulling. Zero drift is fairly low once warmed up, especially compared to the predecessor units like the [[Keithley 610B]] which still relied on electrometer vacuum tubes in the input stage.
The 610C contains only 10 transistors and 2 MOSFETs altogether, which are all used for amplification. The power supply is regulated using Zener diodes.


The 610C contains only 10 transistors and 2 MOS FETs altogether, which are all used for amplification. The power supply is regulated using Zener diodes.
==Links==
* [https://www.youtube.com/watch?v=05zAVuIkNwY Keithley 610C Electrometer teardown] @ YouTube
* [https://www.youtube.com/watch?v=j1bUnSZRRuk Keithley 610C Electrometer Repair] @ YouTube
* [https://www.youtube.com/watch?v=gGYXcRv3Y2w Keithley 610C Electrometer Calibration and Experiments] @ YouTube
* [https://www.radiomuseum.org/r/keithley_solid_state_electrometer_610c.html Keithley 610C Electrometer] @ radiomuseum.org


==Pictures==
==Pictures==
Line 53: Line 74:
Keithley 610C rear.jpg
Keithley 610C rear.jpg
Keithley-610c-interior.jpg
Keithley-610c-interior.jpg
Keithley-610c-input-section.jpg
Keithley-610c-input-section.jpg | Input stage: MOSFETS are thermally coupled, input signal is routed over the top of the circuit board using teflon standoffs
Keithley-610c-range-switch.jpg
Keithley-610c-range-switch.jpg   | Range selector switch with teflon-insulated contacts
Keithley-610c-range-resistors.jpg
Keithley-610c-range-resistors.jpg | Range resistors, from 10 Ω to 10<sup>11</sup> Ω
Keithley-610c-output.jpg
Keithley-610c-output.jpg         | Output stage and power supply
</gallery>
</gallery>


Line 62: Line 83:




[[Category:Voltmeters]]
[[Category:Electrometers]]
[[Category:Keithley]]
[[Category:Keithley]]

Revision as of 07:40, 27 May 2024

Keithley 610C
Solid-state Electrometer
Keithley 610C, frontal view (desktop enclosure, a rack version was also made)

Produced from (?) to (?)

Manuals
(All manuals in PDF format unless noted otherwise)
Manuals – Specifications – Links – Pictures

The Keithley 610C Solid State Electrometer can measure voltages (with a very high input resistance), very low currents, high resistances and charge.

Key Specifications

Voltage 1 mV to 100 V full scale in 11 ranges (×1 / ×3 steps)
Input resistance (volts mode) >1014 Ω or selectable 10 Ω to 1011 Ω in decade steps
Accuracy 1% of full scale, noise 25 μV
Zero offset drift <1 mV per day / 150 μV per °C
Current 10–14 A to 0.3 A full scale in 28 ranges
Resistance 100 Ω to 1014 Ω full scale in 25 linear ranges
Charge 10–13 C to 10–3 C (Amprere-seconds) full scale in 17 ranges
Display Analog meter with mirror scale; positive, negative and center-zero display modes
Power 105−125 V or 210−250 V, 50−60 Hz (some units labelled 50−1000 Hz), 10 W
Accessories
  • Model 2503 Static Detector Probe (for measurements of charge on flat surfaces)
  • Model 6101A Shielded Probe (shielded low-noise cable with a needle-point probe)
  • Model 6103C Divider Probe (1000:1 voltage divider with a 3.8×1011 input resistance)
  • Model 6105 Resistivity Chamber (test fixture for measurement of surface and volume resistivity)

The unit has two sets of outputs on the back panel — one to attach a DVM or recorder, with ±3 V or ±1 mA full scale, using an Amphenol 80-PC2F connector, the other output (marked "×1") tracks the input voltage (in "Normal" mode only, s.b.). It is therefore possible to use the 610C as an amplifier, with a bandwidth of 40 kHz at ×1 down to 100 Hz at maximum amplification.

The input connector is a PL-259 ("UHF") socket. Newer electrometers use triaxial input connectors, with the inner shield (guard) driven by an amplifier to be at the same potential as the input, in order to minimize leakage currents. It is possible to use a guarded input connection on the 610C through an adapter, although the guard potential is only available on the back of the instrument, at the "× 1" output. (The terminal labelled "guard" is at input ground potential.)

Measurement Principle

When set to the "Normal" feedback mode, the 610C measures an input voltage, or an input current as the voltage drop across a shunt resistor. The large range switch selects the shunt resistance in decade steps between 10 Ω and 1011 Ω, or leaves the input open.

Because only the input voltage is measured, the shunts can also be used as defined input resistances in Volts mode (1 / current range). For example, in the 1 nA (10-9 A) range, a 1 GΩ resistance is connected across the input.

Note that in Normal mode, the "×1" output on the rear follows the input voltage within the full range, i.e. when the input is at +100 V, so is that output (for loads ≤100 μA).

Principle of transimpedance amplifier ("Fast" feedback mode in 610C)

In "Fast" feedback mode, the current range resistor is connected between the amplifier output and the input, turning the instrument into a feedback amperemeter (transimpedance amplifier) and reducing the input voltage burden to less than 100 μV. (Modern pico/nanoamperemeters work this way.)

Because there is very little (change of) input voltage, the meter's input capacitance has no effect, and the meter responds faster, hence the name.

In the Coulomb ranges (only valid in Fast mode), a capacitor takes the place of the feedback resistor, the instrument integrates input currents and thereby measures charge in Ampere-seconds (Coulombs).

Internals

At the core of the 610C is an operational amplifier with MOSFET inputs that is constructed from discrete transistors and has a ±100 V range for input and output voltages. This is done by bootstrapping – the output stage operates on ±120 V supplies, and the actual amplifier uses a floating ±9 V supply referred to that output. As a result, the instrument can measure voltages of up to 100 V in either polarity without needing a resistive input divider, thereby preserving the MOSFET gate's high input resistance of >1014 Ω over the entire ±100 V input range.

The opamp has a three-level zero adjustment, with switches for coarse (at the rear panel) and medium levels, and a 10-turn potentiometer for fine nulling. Zero drift is fairly low once warmed up, especially compared to the predecessor units like the Keithley 610B which still relied on electrometer vacuum tubes in the input stage.

The 610C contains only 10 transistors and 2 MOSFETs altogether, which are all used for amplification. The power supply is regulated using Zener diodes.

Links