The '''Tektronix SD-14 High Impedance Probe Sampler''' is a dual-channel, 3 GHz probe sampler designed for high-impedance (100 kΩ) signal acquisition. | The '''Tektronix SD-14 High Impedance Probe Sampler''' is a dual-channel, 3 GHz probe sampler designed for high-impedance (100 kΩ) signal acquisition. |
Latest revision as of 02:58, 24 October 2023
The Tektronix SD-14 High Impedance Probe Sampler is a dual-channel, 3 GHz probe sampler designed for high-impedance (100 kΩ) signal acquisition.
It consists of two samplers, each attached at the end of a 1-meter cable. Applications include high-speed device characterization such as ECL, CMOS, ACL and GaAs testing. Option 01 adds a longer probe tip, two probe-to-board adapters, and two probe-to-SMB adapters. These ease probing of plated holes (as found in Outboards for IC testers) and semi-permanent connections.
In conjunction with the 11801C and SM-11 multi-channel unit, it is possible to configure up to 136 channels of high-impedance acquisition with accuracy approaching 140 ps.
Key Specifications
Rise time | <140 ps |
---|---|
Bandwidth | 3 GHz typical |
Input impedance | 100 kΩ // 0.5 pF |
Dynamic range | 7 V (±3.5 V offset) |
Maximum safe input voltage | ±15 V dc + peak ac |
Time matching range | 350 ps |
Displayed noise | 8 mVRMS max |