The purpose of delay lines in oscilloscopes is to allow observation
of the leading edge of the trigger event.  There is typically a trigger
pick-off, which supplies an undelayed copy of the vertical signal to the the trigger
and sweep circuitry.  Trigger and sweep circuitry needs about 60ns to take action
when presented with the trigger event.  Without a delay line, the trigger event
would already have come and gone before the scope can trigger and sweep.  By sending
the input signal through a delay line, the scope will have triggered and begun sweeping
by the time the trigger event is emerging from the delay line.
The first Tektronix scope to contain a delay line was the [[513|513D]], which uses an L-C network.   
The first Tektronix scope to contain a delay line was the [[513|513D]], which uses an L-C network.   
Soon after, the [[517]] appeared, using a coaxial cable as a delay line.
Soon after, the [[517]] appeared, using a coaxial cable as a delay line.
Line 4: Line 13:
coil of low-loss air-dielectric semi-rigid coax.
coil of low-loss air-dielectric semi-rigid coax.


Some delay lines are at the input
The location of the delay line in the instrument block diagram varies:
Some are between amplifiers
* Delay at input: [[4S1]], [[1S1]]
Some, such as the in the [[516]],
* Delay between amplifiers: [[517]], [[7000 series scopes]]
are between the final amplifier  
* Delay between amplifier and CRT: [[516]], [[551]], [[555]]
and the CRT vertical deflection plates.
 
Delay lines have a similar purpose in sampling scopes.


There are four main ways delay lines are implemented
There are four main ways delay lines are implemented
Line 19: Line 25:
* Helically wound differential transmission line inside braided shield: [[545|545B]], [[547]]
* Helically wound differential transmission line inside braided shield: [[545|545B]], [[547]]
* L-C: [[511|511AD]], [[513|513D]], [[541]], [[545]], [[551]], [[555]]
* L-C: [[511|511AD]], [[513|513D]], [[541]], [[545]], [[551]], [[555]]
The purpose of delay lines in sampling scopes is
the same as for conventional non-sampling scopes. 
Since sampling scopes are often used for observing fast pulses,
delay lines are often problematic since they have dispersion,
and therefore distort pulse waveforms. 
For example, the [[4S1]] contains a delay network
while the [[4S2]] does not.  Because of this,
the 4S2 is a less convenient instrument,
but has faster rise-time the the 4S1. 
One way around the trade-off between pulse response and convenience
is to use random sampling.  Random sampling is provided by the [[3T2]]
and other sampling sweep modules.

Revision as of 20:42, 13 February 2012

The purpose of delay lines in oscilloscopes is to allow observation of the leading edge of the trigger event. There is typically a trigger pick-off, which supplies an undelayed copy of the vertical signal to the the trigger and sweep circuitry. Trigger and sweep circuitry needs about 60ns to take action when presented with the trigger event. Without a delay line, the trigger event would already have come and gone before the scope can trigger and sweep. By sending the input signal through a delay line, the scope will have triggered and begun sweeping by the time the trigger event is emerging from the delay line.

The first Tektronix scope to contain a delay line was the 513D, which uses an L-C network. Soon after, the 517 appeared, using a coaxial cable as a delay line. 7000-series scopes use special twin-lead delay cables. The delay line in the 519 is a large coil of low-loss air-dielectric semi-rigid coax.

The location of the delay line in the instrument block diagram varies:

There are four main ways delay lines are implemented in Tektronix scopes:

  • Conventional coaxial cable: 517, 4S1, 1S1
  • Semi-rigid coax: 519, 113
  • Unshielded helically wound differential transmission line: 581, 585
  • Helically wound differential transmission line inside braided shield: 545B, 547
  • L-C: 511AD, 513D, 541, 545, 551, 555

The purpose of delay lines in sampling scopes is the same as for conventional non-sampling scopes. Since sampling scopes are often used for observing fast pulses, delay lines are often problematic since they have dispersion, and therefore distort pulse waveforms. For example, the 4S1 contains a delay network while the 4S2 does not. Because of this, the 4S2 is a less convenient instrument, but has faster rise-time the the 4S1. One way around the trade-off between pulse response and convenience is to use random sampling. Random sampling is provided by the 3T2 and other sampling sweep modules.