S-2: Difference between revisions
No edit summary |
No edit summary |
||
Line 31: | Line 31: | ||
==Repair issues== | ==Repair issues== | ||
The [[ | The input should always be terminated when not in use because the [[sampling diodes]] are easily damaged by excessive inputs or static discharge. See that article for possible replacements. | ||
==Pictures== | ==Pictures== |
Revision as of 06:33, 17 April 2019
Template:Plugin Sidebar 2 The Tektronix S-2 Sampling Head plug-in provides a single GR-874 50 Ω input with a 75 ps risetime.
The S-2 can be used in the 7000 series and 3S series of sampling plug-ins. It provides an internal trigger pickoff signal for internal triggering. Noise is specified as less than 10 mV.
When the S-2 was first introduced, the rise-time specification was 50 ps, but that was not achieved in practice, so the rise-time specification was revised to be 75 ps.
The interface between the GR-874 connector and the printed circuit board is described in US Patent 3,426,311.
Key Specifications
Rise time | 75 ps |
---|---|
Bandwidth | 4.66 GHz eq. |
Input voltage | −1 to +1 V (max. −5 to +5) |
Repair issues
The input should always be terminated when not in use because the sampling diodes are easily damaged by excessive inputs or static discharge. See that article for possible replacements.
Pictures
-
-
early S-2 with tr ≤ 50ps on front
-
early S-2 with overstickered tr ≤ 50ps
-
two S-2 engineering sample (late model)
-
S-52 and S-2 give 76 ps rise-time together
-
schematic
-
S-2 interior, strobe board side
-
S-2 interior, amplifier board side
-
Sampling diodes in place
-
Input termination resistor
-
Sampler board, right
-
Sampler board, left
-
Diode holder slides out
-
Sampler board underside
-
S-2 with amplifier board removed
-
Strobe board underside
-
S-2 with strobe board removed
-
S-2 pulse response