The remote control connector on the rear of the 7904 scope is an Amphenol 165-16 socket.

Pin Function Notes
A Single sweep reset Close to ground within 0.5V, 10mA max, minimum 10 μs pulse width
B Ground
C A Remote ready +5 V via 47 Ω when active, can drive #49 bulb
D B Remote ready +5 V via 47 Ω when active, can drive #49 bulb
E Remote readout lock-out Close to ground within 0.1V, 2mA max, 2V open voltage
F Remote single-shot lock-out Closure to ground within 0.4V, pin E was grounded before, rate of change ≥ 0.1V/μs, 3mA max, 10V open voltage
H n/c
J n/c
K n/c