PG506: Difference between revisions

From TekWiki
Jump to navigation Jump to search
mNo edit summary
No edit summary
Line 3: Line 3:
* [[Media:070-1740-01.pdf|Tektronix PG506 Manual for sn 39999 and below (OCR, PDF)]]
* [[Media:070-1740-01.pdf|Tektronix PG506 Manual for sn 39999 and below (OCR, PDF)]]
* [http://w140.com/smb/PG506_SM.pdf Tektronix PG506 Manual for sn 40000 and above (OCR, PDF)]
* [http://w140.com/smb/PG506_SM.pdf Tektronix PG506 Manual for sn 40000 and above (OCR, PDF)]
}}
}} In functionally replaces the [[067-0502-01]] in the "standard amplitude" mode, and the [[106]] in the "fast rise" and "high amplitude" modes.


There are (at least) two versions of the PG506. The older version has two relays that switch a low-pass (noise) filter into the 10 V and below ranges. The new version just has a capacitor across the output for these ranges.
There are (at least) two versions of the PG506. The older version has two relays that switch a low-pass (noise) filter into the 10 V and below ranges. The new version just has a capacitor across the output for these ranges.

Revision as of 12:32, 3 October 2019

Tektronix PG506
1 MHz calibration generator
Tektronix PG506

Produced from 1974 to 1993

Manuals
Manuals – Specifications – Links – Pictures

The Tektronix PG506 is a 1 MHz calibration generator plug-in for the TM500 system.

In functionally replaces the 067-0502-01 in the "standard amplitude" mode, and the 106 in the "fast rise" and "high amplitude" modes. 

There are (at least) two versions of the PG506. The older version has two relays that switch a low-pass (noise) filter into the 10 V and below ranges. The new version just has a capacitor across the output for these ranges.

Common failure modes: In case of low or no output on the 20 V and above ranges, check Q270, Q280 and Q290.

Pictures

PG506 older version

PG506 newer version

PG506A