* [[1S1/Repairs|David Wise 1S1 Repair Notes]]
* [[1S1/Repairs|David Wise 1S1 Repair Notes]]
}}
}}
The '''Tektronix 1S1''' is a [[sampling oscilloscope|sampling]] plug-in for [[500-series scopes]].
The '''Tektronix 1S1''' is a [[sampling oscilloscope|sampling]] plug-in for [[500-series scopes]], designed by [[Chuck Edgar]], [[introduced in 1965]] and sold through 1973.


It is essentially one channel of a sampling unit and a timing unit from a [[661]].  It has its own trigger and timebase.   
It is essentially one channel of a sampling unit and a timing unit from a [[661]].  It has its own trigger and timebase.   


The oscilloscope's timebase is usually set to a relatively slow sweep rate when using a 1S1.  Alternatively, the sweep on the oscilloscope can be free-run,  
The oscilloscope's timebase is usually set to a relatively slow sweep rate when using a 1S1.  Alternatively, the sweep  
and the horizontal voltage from the oscilloscope can be used to control the 1S1, as described below.  
on the oscilloscope can free-run, and the horizontal voltage from the oscilloscope can be used to control the 1S1, as described below.  


The 1S1 has a 50 Ω [[GR-874 connector]] for signal input.  The plug-in supplies power (+100 V and −12.6 V) to accessories like the [[P6032]] Cathode Follower Probe, the [[281]] TDR Pulser, and the [[282]] Probe Adapter.
The 1S1 has a 50 Ω [[GR-874 connector]] for the signal input.  The plug-in supplies power (+100 V and −12.6 V) to accessories like  
the [[P6032]] Cathode Follower Probe, the [[281]] TDR Pulser, and the [[282|Type 282 Probe Adapter]], which is an active buffer to
allow the use of high-impedance probes with the 50 Ω input of the 1S1 (at reduced bandwidth).


{{MissingSpecs}}
==Operation==
The 1S1 can operate in normal self-swept mode or an external sweep signal can be applied to the 1S1.   
The 1S1 can operate in normal self-swept mode or an external sweep signal can be applied to the 1S1.   
With the 1S1's internal sweep disabled, the horizontal-in and vertical-out connections can be used  
With the 1S1's internal sweep disabled, the horizontal-in and vertical-out connections can be used  
Line 41: Line 46:
The horizontal-in voltage can also be produced, in "manual" mode, by setting a knob on the 1S1.
The horizontal-in voltage can also be produced, in "manual" mode, by setting a knob on the 1S1.


When the 1S1 is set for INT trigger signal, the EXT TRIG input is terminated with 50 Ω.
==Internals==
The sampler in the 1S1 is quite different from the sampler in the [[1S2]].  The 1S1 uses a [[sampling diodes|four-diode]] sampling bridge, terminated within the 1S1.  The 1S2 uses a two-diode feed-through sampler, typical for TDR instruments.
The sampler in the 1S1 is quite different from the sampler in the [[1S2]].  The 1S1 uses a [[sampling diodes|four-diode]] sampling bridge, terminated within the 1S1.  The 1S2 uses a two-diode feed-through sampler, typical for TDR instruments.
Type 1S1 was designed by [[Chuck Edgar]], [[introduced in 1965]] and sold through 1973.
The [[282|Type 282 Probe Adapter]] works with the 1S1.
It is an active buffer to allow the use of high impedance probes with the 50 Ω input of the 1S1.
It does, however, lower the bandwidth.
When the 1S1 is set for INT trigger signal, the EXT TRIG input is terminated with 50 Ω.


==Pictures==
==Pictures==

Revision as of 01:58, 15 January 2021

Template:Plugin Sidebar 2 The Tektronix 1S1 is a sampling plug-in for 500-series scopes, designed by Chuck Edgar, introduced in 1965 and sold through 1973.

It is essentially one channel of a sampling unit and a timing unit from a 661. It has its own trigger and timebase.

The oscilloscope's timebase is usually set to a relatively slow sweep rate when using a 1S1. Alternatively, the sweep on the oscilloscope can free-run, and the horizontal voltage from the oscilloscope can be used to control the 1S1, as described below.

The 1S1 has a 50 Ω GR-874 connector for the signal input. The plug-in supplies power (+100 V and −12.6 V) to accessories like the P6032 Cathode Follower Probe, the 281 TDR Pulser, and the Type 282 Probe Adapter, which is an active buffer to allow the use of high-impedance probes with the 50 Ω input of the 1S1 (at reduced bandwidth).

Key Specifications

  • please add

Operation

The 1S1 can operate in normal self-swept mode or an external sweep signal can be applied to the 1S1. With the 1S1's internal sweep disabled, the horizontal-in and vertical-out connections can be used so the 1S1 acts as a lookup table, a mapping of X to Y, a function. The 661 also has this capability, in its "A vert/B horiz" mode, which is like X-Y mode for a sampler. In this mode, the horizontal-in voltage controls the time after the trigger event when the sample should be taken, and the vertical-out voltage corresponds to the voltage measured at that instant. This allows a waveform to be digitized using an arbitrarily slow DAC to generate the horizontal voltage and ADC to read the sampled output. But perhaps more importantly, by setting a constant horizontal-in voltage, it allows the output signal at one equivalent time instant to be processed in the time domain. For example, this allows the voltage to be low-pass filtered, so that it can be more accurately measured. The other reason why one might want a signal that consists of multiple sequential measurements of the same equivalent-time instant is that statistics can be calculated on these observations. For example, one might want to know, when testing a logic gate, what is the propagation delay such that 99.9% of the transitions happen faster than this.

The horizontal-in voltage can also be produced, in "manual" mode, by setting a knob on the 1S1.

When the 1S1 is set for INT trigger signal, the EXT TRIG input is terminated with 50 Ω.

Internals

The sampler in the 1S1 is quite different from the sampler in the 1S2. The 1S1 uses a four-diode sampling bridge, terminated within the 1S1. The 1S2 uses a two-diode feed-through sampler, typical for TDR instruments.

Pictures