067-0502-01: Difference between revisions

From TekWiki
Jump to navigation Jump to search
No edit summary
No edit summary
Line 29: Line 29:


{{BeginSpecs}}
{{BeginSpecs}}
{{Spec | Output Signal | DC+, DC-, Squarewave}}
{{Spec | Output | DC+, DC-, Squarewave}}
{{Spec | Amplitude Range | 0.2 mV to 100 V in 1-2-5 sequence, 5mA in current loop}}
{{Spec | Amplitude range | 0.2 mV to 100 V in 1-2-5 sequence, 5mA in current loop}}
{{Spec | Amplitude Accuracy | Within 0.25% into 1 MΩ load}}
{{Spec | Amplitude accuracy | Within 0.25% into 1 MΩ load}}
{{Spec | Squarewave Repetition Rate | 1 kHz within 20% }}
{{Spec | Squarewave repetition rate | 1 kHz within 20% }}
{{Spec | Maximum Input Voltage | 100 V }}
{{Spec | Maximum input voltage | 100 V }}
{{EndSpecs}}
{{EndSpecs}}



Revision as of 09:23, 21 January 2022

Tektronix 067-0502-01
Amplitude calibrator

Produced from (?) to (?)

Manuals
(All manuals in PDF format unless noted otherwise)
Manuals – Specifications – Links – Pictures

The Tektronix 067-0502-01 Amplitude Calibrator and Comparator is an instrument that can be used to produce a reference rectangle voltage or current waveform and/or to compare an external unknown input with an internal or external reference.

The output voltage is only accurate if terminated in 1 MΩ, the 50 V setting will experience the highest error if not properly terminated (0.5% in an open load).

The comparison is performed using an Airpax A175 chopper to switch between the two voltages that are being compared. The resulting output is a rectangular waveform with amplitude equal to the difference of two voltages. This waveform can be amplified using a drift-free AC amplifier and viewed on an oscilloscope for high sensitivity comparison.

See the 067-0502-00 page for a comparison of the -00 and -01 models.

The HV supply uses a tube V427, to regulate the +300V supply. The manual lists V427 as a 7734, a triode-pentode, however some models contain a small circular adapter board (670-1860-00,388-2423-00) to adapt the noval tube socket to a Compactron socket and use a 6G11 double pentode instead. The adapter PCB connects the screen grid and plate of the beam power tube together in a triode configuration. These models also have 6G11 printed below V427 on the chassis.

The output amplitude can be reduced down to 0.2 μV using the 067-0529-00 1000:1 precision divider.

Key Specifications

Output DC+, DC-, Squarewave
Amplitude range 0.2 mV to 100 V in 1-2-5 sequence, 5mA in current loop
Amplitude accuracy Within 0.25% into 1 MΩ load
Squarewave repetition rate 1 kHz within 20%
Maximum input voltage 100 V

Links

Pictures


Some Parts Used in the 067-0502-01

Part Part Number(s) Class Description Used in
6G11 154-0660-00 Vacuum Tube (Pentode) double sharp cut-off beam power pentode 125 132 519 067-0502-01 050-0551-00