* [ | * [[Media:010-6056-03.pdf|Tektronix P6056/P6057 Manual]] | ||
}} | }} | ||
The '''Tektronix P6056''' is a 3.5 GHz ×10 probe for 50 Ω-input scopes, released along with the [[7A19]] amplifier in 1972. | The '''Tektronix P6056''' is a 3.5 GHz ×10 probe for 50 Ω-input scopes, released along with the [[7A19]] amplifier in 1972. |
Revision as of 02:04, 21 July 2022
The Tektronix P6056 is a 3.5 GHz ×10 probe for 50 Ω-input scopes, released along with the 7A19 amplifier in 1972.
It has an Identify switch and readout contact.
Its sister model P6057 has ×100 attenuation.
Key Specifications
Bandwidth | DC to 3.5 GHz |
---|---|
Rise time | 100 ps |
Attenuation | 10× |
Input impedance | 500 Ω // approx. 1 pF |
Scope input impedance | 50 Ω |
Maximum input Voltage | 16 V DC or RMS |
Features | Readout Pin, ID Button |
WARNING:
This high performance device has a fragile resistor assembly in the probe head, subject to failure in rough usage.
Use with slip-on tip whenever possible. One replacement part is provided with the package.
Contact nearest Field Office for additional parts (P/N 307-0330-00), if required.
For replacement and Calibration Procedure, refer to Pg. 3 & 4 of manual.