A6303: Difference between revisions
Jump to navigation
Jump to search
No edit summary |
No edit summary |
||
Line 9: | Line 9: | ||
|discontinued=(?) | |discontinued=(?) | ||
|manuals= | |manuals= | ||
* [[Media:070-3906-01.pdf|Tektronix A6303 Manual]] | * [[Media:070-3906-03.pdf|Tektronix A6303 Manual -03]] | ||
* [[Media:070-3906-01.pdf|Tektronix A6303 Manual -01]] | |||
}} | }} | ||
The '''Tektronix A6303''' aka '''P6303''' is a 15 MHz, 100 A current probe. | The '''Tektronix A6303''' aka '''P6303''' is a 15 MHz, 100 A current probe. |
Revision as of 02:38, 15 October 2022
The Tektronix A6303 aka P6303 is a 15 MHz, 100 A current probe.
It requires an AM503, AM5030 or 11A16 amplifier for operation. The Tektronix 067-0865-00 is a calibration fixture for the P6303.
Key Specifications
Bandwidth | DC − 15 MHz |
---|---|
Rise time | ≤23 ns |
Input range | 100 A continuous, 500 A peak, ≤10 kA · μs (with 11A16: 52.5 A continuous)
Derating: 200 Ap-p up to 20 kHz, down to 20 Ap-p at 20 MHz (see curve) |
Sensitivity | 10 mA/Div to 50 A/Div, 1−2−5 (with AM503) |
Noise | ≤3 mA (@ 100 MHz amplifier BW, 10 mA range, DC) |
Links
- 120-1102-00 Current Transformer
- http://www.tek.com/datasheet/current-measurement-system-probes
- Tek A6303 @ barrytech.com
Pictures
-
P6303 probe
-
Open jaws
-
A6303 maximum current vs frequency
-
A6303 insertion impedance vs frequency