{| class="wikitable" | {| class="wikitable" | ||
! | ! | ||
! P6019 passive | ! P6019 + [[011-0078-00]] passive termination | ||
! P6019 + [[134]] | ! P6019 + [[134]] | ||
! P6020 passive | ! P6020 + [[011-0079-00]] passive termination | ||
! P6020 + [[134]] | ! P6020 + [[134]] | ||
|- | |- | ||
Line 44: | Line 44: | ||
* [http://www.analogdesign.be/vintage-gear/tektronix-p6019-ac-current-probe P6019 @ analogdesign.be] | * [http://www.analogdesign.be/vintage-gear/tektronix-p6019-ac-current-probe P6019 @ analogdesign.be] | ||
==Pictures== | |||
<gallery> | |||
011-0079-00 1.jpg | [[011-0079-00]] passive termination for P6020 | |||
011-0079-00 2.png | [[011-0079-00]] passive termination for P6020 | |||
</gallery> | |||
[[Category:Oscilloscope current probes]] | [[Category:Oscilloscope current probes]] |
Revision as of 09:39, 11 January 2023
The P6019 and P6020 split-core AC current probes were introduced in 1966 together with passive terminations and the Type 134 amplifier.
The probes had no internal frequency compensation components, contrary to their 1969 successor types P6021 and P6022.
Specifications
P6019 + 011-0078-00 passive termination | P6019 + 134 | P6020 + 011-0079-00 passive termination | P6020 + 134 | |
---|---|---|---|---|
Sensitivity | 2 mA/mV or 10 mA/mV | 1 mA/Div to 1 A/Div (scope @ 50 mV/Div) | 1 mA/mV or 10 mA/mV | 1 mA/Div to 1 A/Div (scope @ 50 mV/Div) |
Bandwidth (-3 dB) | 60 MHz @ 10 mA/mV | > 40 MHz | 200 MHz @ 10 mA/mV | > 70 MHz |
LF limit (-3 dB) | 450 Hz @ 2 mA/mV, 120 Hz @ 10 mA/mV | < 12 Hz | 6500 Hz @ 1 mA/mV, 775 Hz @ 10 mA/mV | < 50 Hz |
Maximum current | 15 Ap-p | 6 Ap-p |
Links
Pictures
-
011-0079-00 passive termination for P6020
-
011-0079-00 passive termination for P6020