DM501: Difference between revisions

From TekWiki
Jump to navigation Jump to search
No edit summary
No edit summary
Line 10: Line 10:


{{BeginSpecs}}
{{BeginSpecs}}
{{Spec | DC Voltage ranges and accuracy    | 2 V<sub>DC</sub> to 200 V<sub>DC</sub> in decade ranges (0.1% + 0.01% FS), 1000 V<sub>DC</sub> (0.1% + 0.02% FS) }}
{{Spec | DC Voltage | 2 V<sub>DC</sub> to 200 V<sub>DC</sub> in decade ranges (0.1% + 0.01% FS), 1000 V<sub>DC</sub> (0.1% + 0.02% FS) }}
{{Spec | DC Voltage characteristics        | Input resistance 10 MΩ, CMRR >100 dB at DC, 80 dB at 60 Hz (1 kΩ unbal.), NMRR ≥30 dB at 60 Hz + 20 dB/decade }}
{{Spec | DC Voltage characteristics        | Input resistance 10 MΩ, CMRR >100 dB at DC, 80 dB at 60 Hz (1 kΩ unbal.), NMRR ≥30 dB at 60 Hz + 20 dB/decade }}
{{Spec | AC Voltage ranges and accuracy    | 2 V<sub>DC</sub> to 200 V<sub>AC</sub> in decade ranges, and 500 V<sub>AC</sub> (0.7% + 0.01% FS for sine inputs only) }}
{{Spec | AC Voltage | 2 V<sub>DC</sub> to 200 V<sub>AC</sub> in decade ranges, and 500 V<sub>AC</sub> (0.7% + 0.01% FS for sine inputs only) }}
{{Spec | AC Voltage characteristics        | Input impedance 10 MΩ // ≤70 pF, CMRR unspecified }}
{{Spec | AC Voltage characteristics        | Input impedance 10 MΩ // ≤70 pF, CMRR unspecified }}
{{Spec | DC Current ranges and accuracy    | 2 mA to 2 A in decade ranges, 0.2% + 0.05% FS }}
{{Spec | DC Current | 2 mA to 2 A in decade ranges, 0.2% + 0.05% FS }}
{{Spec | AC Current ranges and accuracy    | 2 mA to 2 A in decade ranges, 0.6% + 0.05% FS (for 40 Hz to 1 kHz sine) }}
{{Spec | AC Current | 2 mA to 2 A in decade ranges, 0.6% + 0.05% FS (for 40 Hz to 1 kHz sine) }}
{{Spec | Ohms ranges and accuracy          | 2 kΩ to 20 MΩ in decade ranges, current 1 mA to 0.1 μA respectively; 0.3% (20 MΩ: 0.5%) + 0.01% FS }}
{{Spec | Ohms | 2 kΩ to 20 MΩ in decade ranges, current 1 mA to 0.1 μA respectively; 0.3% (20 MΩ: 0.5%) + 0.01% FS }}
{{Spec | Temperature range and accuracy    | –55°C to +125°C (±1.5°C), +125°C to 150°C (±2.5°C) }}
{{Spec | Temperature | –55°C to +125°C (±1.5°C), +125°C to 150°C (±2.5°C) }}
{{EndSpecs}}
{{EndSpecs}}


==Options==
==Options==
* Option 1: Odering without P6058 Temp Probe
* Option 1: Ordering without P6058 Temp Probe
* Option 2: Without Temp Capability and Probe (visible by no ''Temp Out'' at the Front)
* Option 2: Without Temp Capability and Probe (visible by no ''Temp Out'' at the Front)



Revision as of 00:26, 6 September 2023

Tektronix DM501
4½ digit multimeter
Tektronix DM501

Produced from 1972 to 1980

Manuals
(All manuals in PDF format unless noted otherwise)
Manuals – Specifications – Links – Pictures

The Tektronix DM501 is a 4½ digit multimeter plug-in for the TM500 system.

(See DM501A for the significantly different successor model.)

It also has a temperature input, requiring the P6058 temperature probe with a 4-pin LEMO S-series connector.

The DM501 MOD718D variant replaced the temperature range with a 200 mV input range providing 10 μV resolution (see Physical Measurements Handbook).

Key Specifications

DC Voltage 2 VDC to 200 VDC in decade ranges (0.1% + 0.01% FS), 1000 VDC (0.1% + 0.02% FS)
DC Voltage characteristics Input resistance 10 MΩ, CMRR >100 dB at DC, 80 dB at 60 Hz (1 kΩ unbal.), NMRR ≥30 dB at 60 Hz + 20 dB/decade
AC Voltage 2 VDC to 200 VAC in decade ranges, and 500 VAC (0.7% + 0.01% FS for sine inputs only)
AC Voltage characteristics Input impedance 10 MΩ // ≤70 pF, CMRR unspecified
DC Current 2 mA to 2 A in decade ranges, 0.2% + 0.05% FS
AC Current 2 mA to 2 A in decade ranges, 0.6% + 0.05% FS (for 40 Hz to 1 kHz sine)
Ohms 2 kΩ to 20 MΩ in decade ranges, current 1 mA to 0.1 μA respectively; 0.3% (20 MΩ: 0.5%) + 0.01% FS
Temperature –55°C to +125°C (±1.5°C), +125°C to 150°C (±2.5°C)

Options

  • Option 1: Ordering without P6058 Temp Probe
  • Option 2: Without Temp Capability and Probe (visible by no Temp Out at the Front)

Internals

The DM501 is precise to 4½ digits, but not accurate to 4½ digits. The combination of a D-to-A converter built from opamps and discretes combined with a noisy high-frequency isolated power supply make for barely achievable 3½ digit accuracy over all ranges. Every TM500 DVM after the 501 (including the DM501A, the DM502, and so on) used integrated A-to-D converters and a mains-frequency isolated power supply.